Literature DB >> 16376444

A technique of using gated-CT images to determine internal target volume (ITV) for fractionated stereotactic lung radiotherapy.

Jian-Yue Jin1, Munther Ajlouni, Qing Chen, Fang-Fang Yin, Benjamin Movsas.   

Abstract

BACKGROUND AND
PURPOSE: To develop and evaluate a technique and procedure of using gated-CT images in combination with PET image to determine the internal target volume (ITV), which could reduce the planning target volume (PTV) with adequate target coverage. PATIENTS AND METHODS: A skin marker-based gating system connected to a regular single slice CT scanner was used for this study. A motion phantom with adjustable motion amplitude was used to evaluate the CT gating system. Specifically, objects of various sizes/shapes, considered as virtual tumors, were placed on the phantom to evaluate the number of phases of gated images required to determine the ITV while taking into account tumor size, shape and motion. A procedure of using gated-CT and PET images to define ITV for patients was developed and was tested in patients enrolled in an IRB approved protocol.
RESULTS: The CT gating system was capable of removing motion artifacts for target motion as large as 3-cm when it was gated at optimal phases. A phantom study showed that two gated-CT scans at the end of expiration and the end of inspiration would be sufficient to determine the ITV for tumor motion less than 1-cm, and another mid-phase scan would be required for tumors with 2-cm motion, especially for small tumors. For patients, the ITV encompassing visible tumors in all sets of gated-CT and regular spiral CT images seemed to be consistent with the target volume determined from PET images. PTV expanded from the ITV with a setup uncertainty margin had less volume than PTVs from spiral CT images with a 10-mm generalized margin or an individualized margin determined at fluoroscopy.
CONCLUSIONS: A technique of determining the ITV using gated-CT images was developed and was clinically implemented successfully for fractionated stereotactic lung radiotherapy.

Entities:  

Mesh:

Year:  2006        PMID: 16376444     DOI: 10.1016/j.radonc.2005.11.012

Source DB:  PubMed          Journal:  Radiother Oncol        ISSN: 0167-8140            Impact factor:   6.280


  11 in total

1.  Use of combined maximum and minimum intensity projections to determine internal target volume in 4-dimensional CT scans for hepatic malignancies.

Authors:  Jin Liu; Jia-Zhou Wang; Jian-Dong Zhao; Zhi-Yong Xu; Guo-Liang Jiang
Journal:  Radiat Oncol       Date:  2012-01-30       Impact factor: 3.481

2.  Motion management strategies and technical issues associated with stereotactic body radiotherapy of thoracic and upper abdominal tumors: A review from NRG oncology.

Authors:  Edward D Brandner; Indrin J Chetty; Tawfik G Giaddui; Ying Xiao; M Saiful Huq
Journal:  Med Phys       Date:  2017-04-20       Impact factor: 4.071

3.  Determination of patient-specific internal gross tumor volumes for lung cancer using four-dimensional computed tomography.

Authors:  Muthuveni Ezhil; Sastry Vedam; Peter Balter; Bum Choi; Dragan Mirkovic; George Starkschall; Joe Y Chang
Journal:  Radiat Oncol       Date:  2009-01-27       Impact factor: 3.481

4.  A contrast-oriented algorithm for FDG-PET-based delineation of tumour volumes for the radiotherapy of lung cancer: derivation from phantom measurements and validation in patient data.

Authors:  Andrea Schaefer; Stephanie Kremp; Dirk Hellwig; Christian Rübe; Carl-Martin Kirsch; Ursula Nestle
Journal:  Eur J Nucl Med Mol Imaging       Date:  2008-07-26       Impact factor: 9.236

5.  Measurement of time delay for a prospectively gated CT simulator.

Authors:  M Goharian; R F H Khan
Journal:  J Med Phys       Date:  2010-04

6.  A study on the influence of breathing phases in intensity-modulated radiotherapy of lung tumours using four-dimensional CT.

Authors:  W C Wu; C L Chan; Y W Wong; J P Cuijpers
Journal:  Br J Radiol       Date:  2009-09-01       Impact factor: 3.039

7.  Incorporating PET information in radiation therapy planning.

Authors:  M Macmanus; T Leong
Journal:  Biomed Imaging Interv J       Date:  2007-01-01

8.  Efficient approach for determining four-dimensional computed tomography-based internal target volume in stereotactic radiotherapy of lung cancer.

Authors:  Seung-Gu Yeo; Eun Seog Kim
Journal:  Radiat Oncol J       Date:  2013-12-31

9.  Margin selection to compensate for loss of target dose coverage due to target motion during external-beam radiation therapy of the lung.

Authors:  W Kyle Foster; Ernest Osei; Rob Barnett
Journal:  J Appl Clin Med Phys       Date:  2015-01-08       Impact factor: 2.102

10.  In-house auto cutoff sensor device for radiotherapy machine to monitor patient movements.

Authors:  S Senthilkumar; V Ramakrishnan
Journal:  J Appl Clin Med Phys       Date:  2008-06-23       Impact factor: 2.102