Evaluation of a diode array for QA measurements on a helical tomotherapy unit.

K M Langen1, S L Meeks, D O Poole, T H Wagner, T R Willoughby, O A Zeidan, P A Kupelian, K J Ruchala, G H Olivera.   

Abstract

A helical tomotherapy system is used in our clinic to deliver intensity-modulated radiation therapy (IMRT) treatments. Since this machine is designed to deliver IMRT treatments, the traditional field flatness requirements are no longer applicable. This allows the unit to operate without a field flatness filter and consequently the 400 mm wide fan beam is highly inhomogeneous in intensity. The shape of this beam profile is mapped during machine commissioning and for quality assurance purposes the shape of the beam profile needs to be monitored. The use of a commercial diode array for quality assurance measurements is investigated. Central axis beam profiles were acquired at different depths using solid water built-up material. These profiles were compared with ion chamber scans taken in a water tank to test the accuracy of the diode array measurements. The sensitivity of the diode array to variations in the beam profile was checked. Over a seven week period, beam profiles were repeatedly measured. The observed variations are compared with those observed with an on-board beam profile monitor. The diode measurements were in agreement with the ion chamber scans. In the high dose, low gradient region the average ratio between the diode and ion chamber readings was 1.000 +/- 0.005 (+/- 1 standard deviation). In the penumbra region the agreement was poorer but all diodes passed the distance to agreement (DTA) requirement of 2 mm. The trend in the beam profile variations that was measured with the diode array device was in agreement with the on-board monitor. While the calculated amount of variation differs between the devices, both were sensitive to subtle variations in the beam profile. The diode array is a valuable tool to quickly and accurately monitor the beam profile on a helical tomotherapy unit.

Entities:  

Mesh:

Substances:

Year:  2005        PMID: 16370429     DOI: 10.1118/1.2089547

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  4 in total

1.  Simulation for improvement of system sensitivity of radiochromic film dosimetry with different band-pass filters and scanner light intensities.

Authors:  Takeshi Kamomae; Yuki Miyabe; Akira Sawada; Osamu Matoba; Manabu Nakata; Shinsuke Yano; Toru Takakura; Takashi Mizowaki; Akio Itoh; Masahiro Hiraoka
Journal:  Radiol Phys Technol       Date:  2011-03-17

2.  Tomotherapy as a tool in image-guided radiation therapy (IGRT): theoretical and technological aspects.

Authors:  S Yartsev; T Kron; J Van Dyk
Journal:  Biomed Imaging Interv J       Date:  2007-01-01

3.  Measurements to predict the time of target replacement of a helical tomotherapy.

Authors:  Severin Kampfer; Stefan Schell; Marciana N Duma; Jan J Wilkens; Peter Kneschaurek
Journal:  J Appl Clin Med Phys       Date:  2011-11-15       Impact factor: 2.102

4.  Stability of the Helical TomoTherapy Hi·Art II detector for treatment beam irradiations.

Authors:  Karin Schombourg; François Bochud; Raphaël Moeckli
Journal:  J Appl Clin Med Phys       Date:  2014-11-08       Impact factor: 2.102

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.