Literature DB >> 16370425

Sampling size in the verification of manufactured-supplied air kerma strengths.

Luis Isaac Ramos1, Rafael Martinez Monge.   

Abstract

Quality control mandate that the air kerma strengths (S(K)) of permanent seeds be verified, this is usually done by statistics inferred from 10% of the seeds. The goal of this paper is to proposed a new sampling method in which the number of seeds to be measured will be set beforehand according to an a priori statistical level of uncertainty. The results are based on the assumption that the S(K) has a normal distribution. To demonstrate this, the S(K) of each of the seeds measured was corrected to ensure that the average S(K) of its sample remained the same. In this process 2030 results were collected and analyzed using a normal plot. In our opinion, the number of seeds sampled should be determined beforehand according to an a priori level of statistical uncertainty.

Mesh:

Year:  2005        PMID: 16370425     DOI: 10.1118/1.2089627

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  2 in total

1.  I-125 seed calibration using the SeedSelectron(®) afterloader: a practical solution to fulfill AAPM-ESTRO recommendations.

Authors:  Jose Perez-Calatayud; Jose Richart; Damián Guirado; Jordi Pérez-García; Silvia Rodríguez; Manuel Santos; Facundo Ballester
Journal:  J Contemp Brachytherapy       Date:  2012-03-30

2.  Assaying multiple (125)I seeds with the well-ionization chamber SourceCheck(4π) 33005 and a new insert.

Authors:  Cristian Candela-Juan; Facundo Ballester; Jose Perez-Calatayud; Javier Vijande
Journal:  J Contemp Brachytherapy       Date:  2015-12-30
  2 in total