Evaluation of an electrolyte analyser for measurement of concentrations of ionized calcium and magnesium in cats.

S Unterer1, B Gerber, T M Glaus, M Hässig, C E Reusch.   

Abstract

The goal of this study was to evaluate the Nova CRT 8 electrolyte analyser for determination of concentrations of ionized calcium (Ca(i)) and magnesium (Mg(i)) in cats, to determine the effects of sample handling and storage and to establish reference ranges. The precision and analytical accuracy of the Nova CRT 8 analyser were good. The concentrations of Ca(i) and Mg(i) were significantly lower in aerobically handled serum samples than in those handled anaerobically. The concentrations of Ca(i) and Mg(i) differed significantly among whole blood, plasma and serum. In anaerobically handled serum, the concentration of Ca(i) was stable for 8 h at 22 degrees C, for 5 days at 4 degrees C and for 1 week at -20 degrees C. The concentration of Mg(i) was stable for 4 h at 22 degrees C but for less than 24 h at 4 degrees C and for less than 1 week at -20 degrees C. In serum from 36 cats, the reference ranges were 1.20-1.35 mmol/L for Ca(i) and 0.47-0.59 mmol/L for Mg(i). The Nova CRT 8 electrolyte analyser is suitable for determination of Ca(i) and Mg(i) concentrations in cats. Anaerobically handled serum samples are recommended and, stored at room temperature, they yield accurate results when analysed within 4 h.

Entities:  

Mesh:

Substances:

Year:  2005        PMID: 16369879     DOI: 10.1007/s11259-005-3301-1

Source DB:  PubMed          Journal:  Vet Res Commun        ISSN: 0165-7380            Impact factor:   2.459


  34 in total

1.  Serum ionized magnesium: comparison of results obtained with three ion-selective analyzers.

Authors:  H J Huijgen; R Sanders; S A Cecco; N N Rehak; G T Sanders; R J Elin
Journal:  Clin Chem Lab Med       Date:  1999-04       Impact factor: 3.694

2.  Effect of storage on measurement of ionized calcium and acid-base variables in equine, bovine, ovine, and canine venous blood.

Authors:  O Szenci; E Brydl; C A Bajcsy
Journal:  J Am Vet Med Assoc       Date:  1991-11-01       Impact factor: 1.936

3.  Evaluation of an ionized magnesium-pH analyzer--NOVA 8.

Authors:  J Thode; B Juul-Jørgensen; M Seibaek; H Elming; E Borresen; R Jordal
Journal:  Scand J Clin Lab Invest       Date:  1998-04       Impact factor: 1.713

4.  Ionized and total magnesium concentrations in blood from dogs with naturally acquired parvoviral enteritis.

Authors:  F A Mann; G D Boon; C C Wagner-Mann; D S Ruben; D P Harrington
Journal:  J Am Vet Med Assoc       Date:  1998-05-01       Impact factor: 1.936

5.  Concentration of ionized calcium in plasma from cats with urethral obstruction.

Authors:  K J Drobatz; D Hughes
Journal:  J Am Vet Med Assoc       Date:  1997-12-01       Impact factor: 1.936

6.  Idiopathic hypercalcemia in cats.

Authors:  A M Midkiff; D J Chew; J F Randolph; S A Center; S P DiBartola
Journal:  J Vet Intern Med       Date:  2000 Nov-Dec       Impact factor: 3.333

7.  Relationship between ionized and total magnesium in serum.

Authors:  W R Külpmann; M Gerlach
Journal:  Scand J Clin Lab Invest Suppl       Date:  1996

Review 8.  More on the measurement of ionized magnesium in whole blood.

Authors:  C Ritter; M Ghahramani; H J Marsoner
Journal:  Scand J Clin Lab Invest Suppl       Date:  1996

9.  Hypercalcemia and calcium oxalate urolithiasis in cats: a report of five cases.

Authors:  H M McClain; J A Barsanti; J W Bartges
Journal:  J Am Anim Hosp Assoc       Date:  1999 Jul-Aug       Impact factor: 1.023

10.  [Ionized calcium and total calcium in the cat].

Authors:  A Deniz; R Mischke
Journal:  Berl Munch Tierarztl Wochenschr       Date:  1995-03       Impact factor: 0.328

View more
  1 in total

1.  Development and validation of a multivariate predictive model to estimate serum ionized calcium concentration from serum biochemical profile results in cats.

Authors:  Natasha Hodgson; Maureen A McMichael; Rosanne E Jepson; Kevin Le Boedec
Journal:  J Vet Intern Med       Date:  2019-09-12       Impact factor: 3.333

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.