Microwave ablation for atrial fibrillation: dose-response curves in the cardioplegia-arrested and beating heart.

Sydney L Gaynor1, Gregory D Byrd, Michael D Diodato, Yosuke Ishii, Anson M Lee, Sandip M Prasad, Jegan Gopal, Richard B Schuessler, Ralph J Damiano.   

Abstract

BACKGROUND: Microwave ablation has been used to replace the traditional incisions used in the surgical treatment of atrial fibrillation. However, dose-response curves have not been established in surgically relevant models. The purpose of this study was to develop dose-response curves for the Flex 10 (Guidant, Inc) microwave device in both the acute cardioplegia-arrested heart and on the beating heart.
METHODS: Twelve domestic pigs (40 to 45 kg) were subjected to microwave ablation in either the arrested (n = 6) or beating heart (n = 6). The cardioplegia-arrested heart was maintained at 10 degrees to 15 degrees C while six atrial endocardial and seven right ventricular epicardial lesions were created in each animal. On the beating heart, six right atrial and seven ventricular epicardial lesions were created. Ablations were performed for 15, 30, 45, 60, 90, 120, and 150 seconds (65 W, 2.45 GHz). The tissue was stained with 2,3,5-triphenyl-tetrazolium chloride, and sectioned at 5-mm intervals. Lesion depth and width were determined from digital micrographs.
RESULTS: Mean atrial wall thickness was 2.8 mm (range, 1 to 8 mm). In the arrested heart, 94% of atrial lesions were transmural at 45 seconds and 100% were transmural at 90 seconds. In the beating heart, only 20% of atrial lesions were transmural despite prolonged ablation times (90 seconds). Ventricular lesion width and depth increased with duration of application, and were similar on the arrested and beating hearts.
CONCLUSIONS: Microwave ablation produces linear dose-response curves. Transmural lesions can be reliably produced on the arrested heart, but not consistently on the beating heart.

Entities:  

Mesh:

Year:  2006        PMID: 16368338     DOI: 10.1016/j.athoracsur.2005.06.062

Source DB:  PubMed          Journal:  Ann Thorac Surg        ISSN: 0003-4975            Impact factor:   4.330


  17 in total

1.  Pulmonary Microwave Ablation Near the Heart: Antenna Positioning Can Mitigate Cardiac Complications in a Porcine Model.

Authors:  George A Carberry; Elisabetta Nocerino; Peter J Mason; Denise J Schwahn; Scott Hetzel; Alyssa M Turnquist; Fred T Lee; Christopher L Brace
Journal:  Radiology       Date:  2016-10-12       Impact factor: 11.105

2.  Surgical and minimally invasive ablation for atrial fibrillation.

Authors:  Ralph J Damiano; Rochus K Voeller
Journal:  Curr Treat Options Cardiovasc Med       Date:  2006-09

3.  A new era in the surgical treatment of atrial fibrillation: the impact of ablation technology and lesion set on procedural efficacy.

Authors:  Spencer J Melby; Andreas Zierer; Marci S Bailey; James L Cox; Jennifer S Lawton; Nabil Munfakh; Traves D Crabtree; Nader Moazami; Charles B Huddleston; Marc R Moon; Ralph J Damiano
Journal:  Ann Surg       Date:  2006-10       Impact factor: 12.969

4.  Microwave Ablation in Mitral Valve Surgery for Atrial Fibrillation (MAMA).

Authors:  Anders Jönsson; Mika Lehto; Henrik Ahn; Ulf Hermansson; Peter Linde; Anders Ahlsson; Juhani Koistinen; Jukka Savola; Pekka Raatikainen; Martti Lepojärvi; Antero Sahlman; Kalervo Werkkala; Lauri Toivonen; Håkan Walfridsson
Journal:  J Atr Fibrillation       Date:  2012-08-20

5.  Fontan conversion with novel direct ablation after childbirth: report of a case.

Authors:  Tsukasa Ozawa; Takeshiro Fujii; Noritsugu Shiono; Satoshi Hamada; Hiroshi Masuhara; Masanori Hara; Yuki Sasaki; Tomoyuki Katayanagi; Katsunori Yoshihara; Yoshifumi Okano; Shinichi Takatsuki; Tsutomu Saji; Nobuya Koyama; Yoshinori Watanabe
Journal:  Surg Today       Date:  2011-10-04       Impact factor: 2.549

Review 6.  The golden age of minimally invasive cardiothoracic surgery: current and future perspectives.

Authors:  Alexander Iribarne; Rachel Easterwood; Edward Y H Chan; Jonathan Yang; Lori Soni; Mark J Russo; Craig R Smith; Michael Argenziano
Journal:  Future Cardiol       Date:  2011-05

7.  Epicardial ablation performance of a novel radiofrequency device on the beating heart in pigs.

Authors:  Yoshiyuki Watanabe; Timo Weimar; Toshinobu Kazui; Urvi Lee; Richard B Schuessler; Ralph J Damiano
Journal:  Ann Thorac Surg       Date:  2014-02       Impact factor: 4.330

8.  Animal studies of epicardial atrial ablation.

Authors:  Richard B Schuessler; Anson M Lee; Spencer J Melby; Rochus K Voeller; Sydney L Gaynor; Shun-Ichiro Sakamoto; Ralph J Damiano
Journal:  Heart Rhythm       Date:  2009-12       Impact factor: 6.343

Review 9.  Surgical ablation devices for atrial fibrillation.

Authors:  Shelly C Lall; Ralph J Damiano
Journal:  J Interv Card Electrophysiol       Date:  2007-12       Impact factor: 1.900

Review 10.  Computational modelling of microwave tumour ablations.

Authors:  Jason Chiang; Peng Wang; Christopher L Brace
Journal:  Int J Hyperthermia       Date:  2013-06       Impact factor: 3.914

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.