Tomonori Unno1, Yuko M Sagesaka, Takami Kakuda.
Abstract
Accurate monitoring of tea catechins in biological samples might provide a means of better evaluation of their benefits. The aim of the present study was to develop a rapid method for extracting tea catechins from human plasma samples with a solid-phase extraction technique and to subsequently measure their concentrations using an HPLC system. A human plasma sample spiked with known concentrations of the analyte standards was passed through a Waters Oasis HLB cartridge. After repeated washing, tea catechins were eluted with 70% dimethylformamide containing 0.1% phosphoric acid, and the resulting eluate was injected into an HPLC system. Analytes were separated on a reverse-phase C18 column using an isocratic mobile phase and detected electrochemically. The coefficient of variation for inter- and intraday reproducibility was less than 5.0% and 6.4%, respectively. Linearity was established for the concentration range of 0.01-1.0 microM. The method was successfully applied to measure tea catechin concentrations in the plasma of two healthy subjects who received a single ingestion of a green tea beverage. The proposed method enables the rapid and accurate quantitation of plasma tea catechins and might prove useful for the evaluation of beneficial health effects of tea consumption.Entities: