Load-deflection characteristics of nickel-titanium alloy wires after clinical recycling and dry heat sterilization.

S Kapila1, J W Haugen, L G Watanabe.   

Abstract

The desirable mechanical properties of nickel-titanium alloy wires and their relatively high cost has prompted many clinicians to recycle these wires. Clinical recycling exposes the wires to several weeks or months of mechanical stresses and elements of the oral environment, as well as sterilization between uses. In a previous study it was noted that clinical recycling combined with cold sterilization alters the load-deflection characteristics and surface topography of nickel-titanium wires. Whether similar changes in mechanical properties occur in wires subjected to repeated clinical use and dry heat sterilization is not yet clear. The purpose of this investigation was to determine the effects of in vivo recycling interposed by dry heat sterilization (together referred to as clinical recycling, CR) on the load-deflection characteristics of nickel-titanium alloy wires. To differentiate the effects of dry heat sterilization (DHS) from those of CR on the behavior of these wires, a series of wires were also subjected to DHS only. Two types of nickel-titanium wires, namely Nitinol and NiTi were subjected to a three-point bending test in an as-received condition (T0), after one cycle (T1), and two cycles (T2) of DHS or CR. Ten wires comprised each subsample. Statistical analyses were done by one-factor repeated measures analysis of variance and Scheffe F test. Both DHS alone, as well as CR, produced significant changes in the loading and unloading characteristics of Nitinol and NiTi wires. However, the changes in the load-deflection characteristics of these wires after DHS only were relatively small, and the clinical significance of these changes is open to question.(ABSTRACT TRUNCATED AT 250 WORDS)

Entities:  

Mesh:

Substances:

Year:  1992        PMID: 1636628     DOI: 10.1016/0889-5406(92)70023-4

Source DB:  PubMed          Journal:  Am J Orthod Dentofacial Orthop        ISSN: 0889-5406            Impact factor:   2.650


  9 in total

1.  Evaluation of Mechanical and Physical Properties of Clinically Used and Recycled Superelastic NiTi Wires.

Authors:  Venkata Naidu Bavikati; Gowri Sankar Singaraju; Prasad Mandava; Sai Sandeep Killamsetty; Venkatesh Nettam; Praveen Kumar Reddy Karnati
Journal:  J Clin Diagn Res       Date:  2016-07-01

2.  Comparison of Superelasticity of Nickel Titanium Orthodontic Arch wires using Mechanical Tensile Testing and Correlating with Electrical Resistivity.

Authors:  Aravind Sivaraj
Journal:  J Int Oral Health       Date:  2013-06-23

3.  Pain and discomfort experienced after placement of a conventional or a superelastic NiTi aligning archwire. A randomized clinical trial.

Authors:  L M Fernandes; B Ogaard; L Skoglund
Journal:  J Orofac Orthop       Date:  1998       Impact factor: 1.938

4.  Torsional superelasticity of NiTi archwires.

Authors:  Yves Bolender; Anne Vernière; Christophe Rapin; Marie-Pierryle Filleul
Journal:  Angle Orthod       Date:  2010-11       Impact factor: 2.079

5.  Effects of repeated sterilization cycles on primary stability of orthodontic mini-screws.

Authors:  Sercan Akyalcin; Holly P McIver; Jeryl D English; Joe C Ontiveros; Ron L Gallerano
Journal:  Angle Orthod       Date:  2012-12-17       Impact factor: 2.079

6.  A clinical comparison of three aligning archwires in terms of alignment efficiency: A prospective clinical trial.

Authors:  Reem Sh Abdelrahman; Kazem S Al-Nimri; Emad F Al Maaitah
Journal:  Angle Orthod       Date:  2014-08-04       Impact factor: 2.079

7.  Effect of dry heat and steam sterilization on load-deflection characteristics of β-titanium wires: An in vitro study.

Authors:  Shiva Alavi; Neda Sinaee
Journal:  Dent Res J (Isfahan)       Date:  2012-09

8.  Effect of Recycling and Autoclave Sterilization on the Unloading Forces of NiTi Closed-Coil Springs: An In Vitro Study.

Authors:  Sh Momeni Danaei; M Oshagh; A Khozaei
Journal:  J Dent (Shiraz)       Date:  2013-12

9.  Evaluation of tensile strength and surface topography of orthodontic wires after infection control procedures: An in vitro study.

Authors:  M Brindha; N Kurunji Kumaran; K Rajasigamani
Journal:  J Pharm Bioallied Sci       Date:  2014-07
  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.