T Shinozaki1, J Sato, H Watanabe, K Takagishi, J Aoki, Y Koyama, A Takahashi.
Abstract
The nidi of osteoid osteoma are small and difficult to identify precisely; this can necessitate substantial resection of the surrounding normal bone. We applied percutaneous, image-guided radiofrequency ablation to treat 5 patients with osteoid osteoma. The precise location of each nidus was determined using a 3-mm-thick spiral computed tomographic scan. The most appropriate needle pathway was selected to avoid major neural and vascular structures. A 14-gauge core needle was introduced into the nidus and then exchanged for a 17-gauge straight rigid electrode with a 1-cm exposed tip to heat the lesion to 90 degrees centigrade for 5 minutes. Four of the 5 patients were discharged on the day after surgery without any external supports, and were pain-free after a few days. The remaining patient had a pes equinus contracture induced by the penetration of the calf muscles through a posterior approach but was fully recovered 6 days later. The mean follow-up period was 20.6 months. No recurrences or late complications were observed in this series. This case series shows that the computed tomography-guided percutaneous radiofrequency ablation is a simple, minimally invasive, and highly effective technique for the treatment of osteoid osteoma.Entities:
Mesh:
Year: 2005 PMID: 16365501 DOI: 10.1177/230949900501300320