von Willebrand factor ristocetin cofactor (VWF:RCo) assay: implementation on an automated coagulometer (ACL).

R Redaelli1, A R Corno, L Borroni, G Mostarda, M Nichelatti, E Morra, F Baudo.   

Abstract

BACKGROUND: VWF:RCo assay is the standard and widely used laboratory test for von Willebrand disease (VWD) diagnosis. It is hampered by high intra- and inter-assay imprecision and is time consuming. Automation may improve the assay performance and allow its routine application.
OBJECTIVE: Automation of VWF:RCo on the ACL 7000 coagulometer (Instrumentation Laboratory, Milan, Italy) and its evaluation in VWD diagnosis. METHODS AND MATERIALS: Method performance determination: precision, detection limit (DL), interferences, dose-response curve. Method comparison: analysis of 105 plasma samples from normal subjects (50), VWD type 1 (24), VWD type 2 (24) and VWD type 3 (7) with ACL VWF:RCo and comparison with the reference aggregometric (AGM) method.
RESULTS: ACL VWF:RCo: CVs around 10% vs. 19% of AGM method; DL: 0.08 U mL(-1); potential interferences from bilirubin, triglycerides and hemoglobin, avoided by suitable plasma dilution; high correlation with AGM VWF:RCo (Deming regression Y =-0.0277 + 1.0519X) either in normal or VWD plasmas. In VWD types 1 and 2, the VWF:RCo/VWF:Ag ratios are >0.6 or <0.6, respectively, when calculated with both AGM and ACL VWF:RCo values.
CONCLUSIONS: The automated VWF:RCo on the ACL 7000 coagulometer shows precision improvement and high correlation with the reference AGM method. The test allows the diagnosis of both quantitative (VWD types 1 and 3) and qualitative (VWD type 2) forms of the disease. These results and the assay feasibility make it a suitable and reliable test for the routine diagnosis of VWD.

Entities:  

Mesh:

Substances:

Year:  2005        PMID: 16359507     DOI: 10.1111/j.1538-7836.2005.01628.x

Source DB:  PubMed          Journal:  J Thromb Haemost        ISSN: 1538-7836            Impact factor:   5.824


  4 in total

Review 1.  Platelet-dependent von Willebrand factor activity. Nomenclature and methodology: communication from the SSC of the ISTH.

Authors:  I Bodó; J Eikenboom; R Montgomery; J Patzke; R Schneppenheim; J Di Paola
Journal:  J Thromb Haemost       Date:  2015-05-09       Impact factor: 5.824

2.  Optimization of the automated, CS-2000i™ method for measuring low levels of von Willebrand factor ristocetin cofactor activity (VWF:RCo).

Authors:  Tomoko Matsumoto; Keiji Nogami; Masahiro Okuda; Midori Shima
Journal:  Int J Hematol       Date:  2014-12-06       Impact factor: 2.490

3.  A Comparative Evaluation of an Automated Functional Assay for Von Willebrand Factor Activity in Type 1 Von Willebrand Disease.

Authors:  Shiue-Wei Lai; Chia-Yau Chang; Shin-Nan Cheng; Shu-Hsia Hu; Chung-Yu Lai; Yeu-Chin Chen
Journal:  Int J Gen Med       Date:  2021-09-02

Review 4.  Endothelium, Platelets, and Coagulation Factors as the Three Vital Components for Diagnosing Bleeding Disorders: A Simplified Perspective with Clinical Relevance.

Authors:  Abhinav Bhattarai; Sangam Shah; Sara Bagherieh; Omid Mirmosayyeb; Sangharsha Thapa; Sandip Paudel; Pawan Gyawali; Pitambar Khanal
Journal:  Int J Clin Pract       Date:  2022-08-27       Impact factor: 3.149

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.