Determination of theophylline concentration in serum by chemiluminescent immunoassay.

Mei-xia Zhou1, Cha-ying Guan, Guang Chen, Xin-you Xie, Sheng-hai Wu.   

Abstract

OBJECTIVE: This study aimed to establish chemiluminescent immunoassay (CLIA) for quantitative determination of theophylline levels in human serum.
METHODS: To measure the concentration of theophylline (n=122) and evaluate the assay.
RESULTS: The linear range of the CLIA method was 0.51-40 mg/L (Y=1.02X+0.44, r=0.995). The intra and inter CV (coefficient variance) of CLIA were 3.20% and 3.57%, respectively. The average recovery rate was 102.3%. This method was free from interference by brilirubin (<200 micromol/L), hemoglobin (<10 g/L), and triglycerides (<15 mmol/L).
CONCLUSION: This method is simple, convenient and precise for clinical pharmacokinetics study of theophylline.

Entities:  

Mesh:

Substances:

Year:  2005        PMID: 16358370      PMCID: PMC1390635          DOI: 10.1631/jzus.2005.B1148

Source DB:  PubMed          Journal:  J Zhejiang Univ Sci B        ISSN: 1673-1581            Impact factor:   3.066


  15 in total

1.  Comparison of serum digoxin concentration monitoring by fluorescence polarization immunoassay on the TDxFLx and dry chemistry enzyme immunoassay on the Vitros 950.

Authors:  Bogdan Solnica
Journal:  Clin Chem Lab Med       Date:  2004       Impact factor: 3.694

2.  Chemiluminescence immunoassay for chloramphenicol.

Authors:  Si Lin; Shi-Quan Han; Yi-Bing Liu; Wen-Ge Xu; Guo-Ying Guan
Journal:  Anal Bioanal Chem       Date:  2005-06-24       Impact factor: 4.142

3.  An ultraviolet spectrophotometric method for the determination of theophylline and theobromine in blood and tissues.

Authors:  J A SCHACK; S H WAXLER
Journal:  J Pharmacol Exp Ther       Date:  1949-11       Impact factor: 4.030

4.  Detection of antibodies to S. enteritidis in broilers by means of indirect ELISA and chemiluminescent immunoassay (CLIA).

Authors:  B M Zamora; M Hartung; G Hildebrandt; A Käsbohrer
Journal:  Zentralbl Veterinarmed B       Date:  1999-02

5.  [Effect of low dose of inhaled corticosteroid combined with small dose of oral theophylline on treatment of bronchial asthma].

Authors:  J Li; H Mo; H Huang
Journal:  Zhonghua Jie He He Hu Xi Za Zhi       Date:  2000-06

Review 6.  Therapeutic drug monitoring in drug overdose.

Authors:  A H Dawson; I M Whyte
Journal:  Br J Clin Pharmacol       Date:  1999-09       Impact factor: 4.335

7.  Rapid detection of oleander poisoning using digoxin immunoassays: comparison of five assays.

Authors:  Amitava Dasgupta; Pradip Datta
Journal:  Ther Drug Monit       Date:  2004-12       Impact factor: 3.681

8.  Determination of beta2-microglobulin in biological samples using an immunoenzymometric assay (chemiluminescence detection) or an immunoturbidimetric assay: comparison with a radioimmunoassay.

Authors:  N Terrier; A Bonardet; B Descomps; J P Cristol; A M Dupuy
Journal:  Clin Lab       Date:  2004       Impact factor: 1.138

9.  [Clinical value comparison between chemiluminescent immunoassay and immunoradiometric assay in detecting TSH in serum].

Authors:  Jian-ying Yu; Ying Lin
Journal:  Hunan Yi Ke Da Xue Xue Bao       Date:  2003-06

10.  A competitive chemiluminescent immunoassay for the sensitive measurement of apolipoprotein B100.

Authors:  G Ogbonna; P S Caines; P Catomeris; R J Thibert; K Adeli
Journal:  Clin Biochem       Date:  1995-04       Impact factor: 3.281

View more
  3 in total

1.  Development of an electrochemical immunoassay for detection of gatifloxacin in swine urine.

Authors:  Jian Yi; Meng Meng; Zhong-qiu Liu; Jin-fang Zhi; Yuan-yang Zhang; Jing Xu; Ya-bin Wang; Jin-ting Liu; Ri-mo Xi
Journal:  J Zhejiang Univ Sci B       Date:  2012-02       Impact factor: 3.066

Review 2.  Nano optical and electrochemical sensors and biosensors for detection of narrow therapeutic index drugs.

Authors:  Omid Heydari Shayesteh; Reza Mahjub; Akram Ranjbar; Katayoun Derakhshandeh; Mahdi Jamshidi
Journal:  Mikrochim Acta       Date:  2021-11-06       Impact factor: 5.833

3.  A Disposable Sensor Chip Using a Paste Electrode with Surface-Imprinted Graphite Particles for Rapid and Reagentless Monitoring of Theophylline.

Authors:  Tomoji Ohishi; Yasuo Yoshimi
Journal:  Molecules       Date:  2022-04-11       Impact factor: 4.927

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.