| Literature DB >> 16352437 |
D J Kenny1, J M Brown2, M E Palmer2, M F Snel2, R H Bateman2.
Abstract
We present a novel enhancement to matrix-assisted laser desorption ionization (MALDI) post-source decay (PSD) analysis whereby fragment ions from multiple precursor ions are acquired into the same spectrum without employing a timed ion gate to preselect each parent ion. Fragment ions are matched to their corresponding precursor ions by comparing spectra acquired at slightly different reflectron electric fields. By measuring the difference in time-of-flight (TOF) between the two spectra for each fragment, it is possible to calculate the mass of the fragment ion and its parent. This new "parallel PSD" technique reduces analysis time and consumes less sample than conventional PSD, which requires an ion gate for serial preselection of precursor ions.Entities:
Mesh:
Substances:
Year: 2005 PMID: 16352437 DOI: 10.1016/j.jasms.2005.09.006
Source DB: PubMed Journal: J Am Soc Mass Spectrom ISSN: 1044-0305 Impact factor: 3.109