| Literature DB >> 16333604 |
Q L Zhang1, J J Xu, H Z Lian, X Y Li, H Y Chen.
Abstract
A method has been developed for rapidly separating and detecting strychnine and brucine using a poly(dimethysiloxane) (PDMS) microchip and electrochemical (EC) detection. PDMS microchannels dynamically modified by Brij35 are shown to be more efficient than native ones. The two analytes are well separated within 90 s in 70 mmol/L acetate buffer (pH 5.5) containing 0.01% (v/v) Brij35. Detection limits were found to be 1.0 micromol/L for strychnine and 0.2 micromol/L for brucine at S/N = 3. The method was used to determine trace strychnine and brucine in rat serum, and the results obtained correlate well with those obtained via high-performance liquid chromatography (HPLC).Entities:
Mesh:
Substances:
Year: 2005 PMID: 16333604 DOI: 10.1007/s00216-005-0146-y
Source DB: PubMed Journal: Anal Bioanal Chem ISSN: 1618-2642 Impact factor: 4.142