Literature DB >> 16333604

Rapid separation of strychnine and brucine on a dynamically modified poly(dimethylsiloxane) microchip followed by electrochemical detection.

Q L Zhang1, J J Xu, H Z Lian, X Y Li, H Y Chen.   

Abstract

A method has been developed for rapidly separating and detecting strychnine and brucine using a poly(dimethysiloxane) (PDMS) microchip and electrochemical (EC) detection. PDMS microchannels dynamically modified by Brij35 are shown to be more efficient than native ones. The two analytes are well separated within 90 s in 70 mmol/L acetate buffer (pH 5.5) containing 0.01% (v/v) Brij35. Detection limits were found to be 1.0 micromol/L for strychnine and 0.2 micromol/L for brucine at S/N = 3. The method was used to determine trace strychnine and brucine in rat serum, and the results obtained correlate well with those obtained via high-performance liquid chromatography (HPLC).

Entities:  

Mesh:

Substances:

Year:  2005        PMID: 16333604     DOI: 10.1007/s00216-005-0146-y

Source DB:  PubMed          Journal:  Anal Bioanal Chem        ISSN: 1618-2642            Impact factor:   4.142


  1 in total

1.  The Production and Evaluation of an Electrochemical Sensors for Strychnine and Its Main Metabolite Strychnine N-Oxide for Their Use in Biological Samples.

Authors:  Bakhtiyar Qader; Issam Hussain; Mark Baron; Rafael Estevez-Brito; John Paul Cassella; Jose Gonzalez-Rodriguez
Journal:  Molecules       Date:  2022-03-11       Impact factor: 4.411