The accuracy of pulse oximetry at two haematocrit levels.

M Vegfors1, L G Lindberg, P A Oberg, C Lennmarken.   

Abstract

The object of this study was to investigate the influence of haematocrit on the accuracy of pulse oximetry. Seven Swedish land race rabbits were studied. The oxygen saturation of haemoglobin was decreased step-wise using increasing fractions of nitrogen to the inspiratory gas. One pulse oximeter probe was attached on the front leg and another probe directly over the common carotid artery. The pulse oximeter readings (SpO2) were compared with simultaneous oxygen saturation analysis (SaO2) by a haemoximeter. The pulse oximeter measurements were performed at the haematocrit levels of approximately 40% and 11%, respectively. We found a good correlation between SpO2 and SaO2 in a wide range of the oxygen saturation, i.e. SaO2 26-100%. After haemodilution the correlation was improved in the range 86-100%, but not in the range 26-85%. No correlation between SpO2 and SaO2 was found when the sensor was attached directly over the artery during normal haematocrit levels. After haemodilution a better correlation was however obtained. These results indicate that the accuracy of pulse oximetry is dependent on the haematocrit level.

Entities:  

Mesh:

Substances:

Year:  1992        PMID: 1632169     DOI: 10.1111/j.1399-6576.1992.tb03496.x

Source DB:  PubMed          Journal:  Acta Anaesthesiol Scand        ISSN: 0001-5172            Impact factor:   2.105


  1 in total

1.  Pulse oximeter signal at various blood flow conditions in an in vitro model.

Authors:  L G Lindberg; M Vegfors; C Lennmarken; P A Oberg
Journal:  Med Biol Eng Comput       Date:  1995-01       Impact factor: 2.602

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.