Advantages of terminating Zippy Estimation by Sequential Testing (ZEST) with dynamic criteria for white-on-white perimetry.

Allison M McKendrick1, Andrew Turpin.   

Abstract

PURPOSE: A number of automated perimeters use the Zippy Estimation by Sequential Testing (ZEST) algorithm, which is an adaptive Bayesian method, for determining sensitivity measures. There are two popular rules for deciding when to terminate Bayesian procedures: (1) after a fixed number of presentations; or (2) when the probability density function (pdf) over all thresholds modified by the procedure becomes sufficiently narrow (a dynamic termination criterion). It has recently been argued that fixed termination criteria perform equally as well as dynamic criteria when applied in a fashion typical of laboratory-based visual psychophysics. Perimetry, however, has specific requirements; the tests must be very short, there is a wide range of possible sensitivities, and erroneous responses from the patient must be tolerated. This study used computer simulation to compare fixed and dynamic termination criteria for the ZEST algorithm using conditions typical of white-on-white perimetry.
METHODS: Eight ZEST procedures were compared using the following termination criteria: fixed termination after 4, 5, 6, 7, and 8 presentations; dynamic termination when the standard deviation of the pdf was 1 dB, 1.5 dB, and 2 dB. Four patient error models were used: ideal, typical false-positive, typical false-negative, and unreliable patients. We also ran a version of ZEST that set the likelihood function exactly equal to the patient's frequency of seeing curve.
RESULTS: The mean absolute error and standard deviation of error in threshold measurement was higher for the fixed termination criteria than for dynamic termination criteria of the same average number of presentations.
CONCLUSIONS: The results of our simulations indicate that dynamic procedures have some distinct benefits over fixed termination procedures when a minimum of presentations are required and response errors are made as in a white-on-white perimetric setting. Dynamic termination criteria are at least partially successful in expending more presentations when required to enhance test precision.

Entities:  

Mesh:

Year:  2005        PMID: 16317375

Source DB:  PubMed          Journal:  Optom Vis Sci        ISSN: 1040-5488            Impact factor:   1.973


  10 in total

1.  Visual contrast sensitivity in Alzheimer's disease, mild cognitive impairment, and older adults with cognitive complaints.

Authors:  Shannon L Risacher; Darrell Wudunn; Susan M Pepin; Tamiko R MaGee; Brenna C McDonald; Laura A Flashman; Heather A Wishart; Heather S Pixley; Laura A Rabin; Nadia Paré; Jessica J Englert; Eben Schwartz; Joshua R Curtain; John D West; Darren P O'Neill; Robert B Santulli; Richard W Newman; Andrew J Saykin
Journal:  Neurobiol Aging       Date:  2012-10-18       Impact factor: 4.673

2.  Improving Visual Field Examination of the Macula Using Structural Information.

Authors:  Giovanni Montesano; Luca M Rossetti; Davide Allegrini; Mario R Romano; David P Crabb
Journal:  Transl Vis Sci Technol       Date:  2018-12-28       Impact factor: 3.283

3.  Visual field loss in schizophrenia: evaluation of magnocellular pathway dysfunction in schizophrenic patients and their parents.

Authors:  Carolina Pelegrini Barbosa Gracitelli; Fabiana Benites Vaz de Lima; Rodrigo A Bressan; Augusto Paranhos Junior
Journal:  Clin Ophthalmol       Date:  2013-05-29

4.  Increasing the Spatial Resolution of Visual Field Tests Without Increasing Test Duration: An Evaluation of ARREST.

Authors:  Vasanth Muthusamy; Andrew Turpin; Mark J Walland; Bao N Nguyen; Allison M McKendrick
Journal:  Transl Vis Sci Technol       Date:  2020-12-16       Impact factor: 3.283

5.  Towards Patient-Tailored Perimetry: Automated Perimetry Can Be Improved by Seeding Procedures With Patient-Specific Structural Information.

Authors:  Jonathan Denniss; Allison M McKendrick; Andrew Turpin
Journal:  Transl Vis Sci Technol       Date:  2013-05-31       Impact factor: 3.283

6.  Incorporating Spatial Models in Visual Field Test Procedures.

Authors:  Nikki J Rubinstein; Allison M McKendrick; Andrew Turpin
Journal:  Transl Vis Sci Technol       Date:  2016-03-11       Impact factor: 3.283

7.  Usefulness of 10-2 Matrix Frequency Doubling Technology Perimetry for Detecting Central Visual Field Defects in Preperimetric Glaucoma Patients.

Authors:  Younhea Jung; Hae-Young L Park; Yi Ryeung Park; Chan Kee Park
Journal:  Sci Rep       Date:  2017-11-07       Impact factor: 4.379

8.  The Human Touch: Using a Webcam to Autonomously Monitor Compliance During Visual Field Assessments.

Authors:  Pete R Jones; Giorgia Demaria; Iris Tigchelaar; Daniel S Asfaw; David F Edgar; Peter Campbell; Tamsin Callaghan; David P Crabb
Journal:  Transl Vis Sci Technol       Date:  2020-07-20       Impact factor: 3.283

9.  An Open-source Static Threshold Perimetry Test Using Remote Eye-tracking (Eyecatcher): Description, Validation, and Preliminary Normative Data.

Authors:  Pete R Jones
Journal:  Transl Vis Sci Technol       Date:  2020-07-13       Impact factor: 3.283

10.  Effect of Sensitivity Disparity Between the Two Eyes on Pointwise Monocular Sensitivity Under Binocular Viewing in Patients With Glaucoma.

Authors:  Akemi Wakayama; Hiroki Nomoto; Yasutaka Chiba; Chota Matsumoto; Shunji Kusaka
Journal:  J Glaucoma       Date:  2021-01-01       Impact factor: 2.290

  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.