Effects of lung time constant, gas analyser delay and rise time on measurements of respiratory dead-space.

Yongquan Tang1, Martin J Turner, A Barry Baker.   

Abstract

This study evaluated effects of mechanical time constants (tau(m)) of the respiratory system, delays between flow and CO(2) partial pressure (P(CO)(2)) signals and rise time of the CO(2) analyser on dead-space measurements. A computer model simulated low alveolar dead-space, high alveolar dead-space, 0.2 <or= tau(m) <or= 3.6 s and varying ventilation-perfusion ratios (V/Q). CO(2) expirograms were recorded from the model under each condition and from 22 anaesthetized intubated patients. P(CO)(2) signals were shifted with respect to flow to produce varying time delays and anatomic and physiological dead-spaces were calculated. The CO(2) analyser was simulated as a critically damped second-order system with 10-90% rise times of 25-400 ms. The error in measured dead-space increases approximately 2.5% per 10 ms signal delay for normal lungs (tau(m) = 1 s), but has low sensitivity (0.58% per 10 ms) to the rise time of the CO(2) analyser. Sensitivity of physiological dead-space, but not anatomic dead-space to delay is decreased in high alveolar dead-space and abnormal V/Q distribution. Shorter tau(m) increase the error sensitivity of both physiological and anatomic dead-spaces to both delay and rise time. P(CO)(2) and flow should be well synchronized, particularly when tau(m) are short, to avoid dead-space errors.

Entities:  

Mesh:

Year:  2005        PMID: 16311457     DOI: 10.1088/0967-3334/26/6/019

Source DB:  PubMed          Journal:  Physiol Meas        ISSN: 0967-3334            Impact factor:   2.833


  4 in total

1.  Hypoventilation patterns during bronchoscopic sedation and their clinical relevance based on capnographic and respiratory impedance analysis.

Authors:  Yu-Lun Lo; Hau-Tieng Wu; Yu-Ting Lin; Han-Pin Kuo; Ting-Yu Lin
Journal:  J Clin Monit Comput       Date:  2019-02-06       Impact factor: 2.502

2.  Impact of Software Settings on Multiple-Breath Washout Outcomes.

Authors:  Selina Summermatter; Florian Singer; Philipp Latzin; Sophie Yammine
Journal:  PLoS One       Date:  2015-07-13       Impact factor: 3.240

3.  Enhanced photoacoustic gas analyser response time and impact on accuracy at fast ventilation rates during multiple breath washout.

Authors:  Alex Horsley; Kenneth Macleod; Ruchi Gupta; Nick Goddard; Nicholas Bell
Journal:  PLoS One       Date:  2014-06-03       Impact factor: 3.240

Review 4.  Current methodological and technical limitations of time and volumetric capnography in newborns.

Authors:  Gerd Schmalisch
Journal:  Biomed Eng Online       Date:  2016-08-30       Impact factor: 2.819

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.