| Literature DB >> 16309979 |
Sun Ok Choi1, So Young Um, Sung Hee Jung, Seo Jeong Jung, Joo Il Kim, Hwa Jeong Lee, Soo Youn Chung.
Abstract
A direct injection column-switching high-performance liquid chromatography (HPLC) method was developed and validated for quantification of zaltoprofen in rat plasma. Following dilution with mobile phase A, i.e. acetonitrile-10mM potassium phosphate buffer (pH 6.8) (12:88, v/v) samples were directly injected to the pre-column without sample pre-purification step. After endogenous plasma components were eluted to waste, the system was switched and the analyte was eluted to the trap column. Zaltoprofen was then back-flushed to the analytical column for separation with mobile phase B, i.e. acetonitrile-10mM potassium phosphate buffer (pH 6.8) (35:65, v/v) and quantification with an ultraviolet detector at 230 nm. The calibration curve was linear in the concentration range of 40-5000 ngmL(-1). This method has been fully validated and shown to be specific, accurate and precise. The method is simple, rapid and the sample preparation is minimal and appears to be useful for the pharmacokinetic study of zaltoprofen.Entities:
Mesh:
Substances:
Year: 2005 PMID: 16309979 DOI: 10.1016/j.jchromb.2005.11.003
Source DB: PubMed Journal: J Chromatogr B Analyt Technol Biomed Life Sci ISSN: 1570-0232 Impact factor: 3.205