A comparison of device measures of dorsal first ray mobility.

Ward M Glasoe1, Brett R Grebing, Susan Beck, Michael J Coughlin, Charles L Saltzman.   

Abstract

BACKGROUND: Devices built by Glasoe and Klaue have been used in several studies to measure first ray mobility. Both devices measure sagittal motion of the first ray in a dorsal direction. The primary difference in the devices is the method of the load imposed. This study investigates whether first ray mobility measured with the Glasoe device is similar to the amount of mobility measured with the Klaue device.
METHODS: Using the devices described by Glasoe and Klaue, dorsal first ray mobility was measured in 39 patients who had foot and ankle problems. Paired t-tests were computed to assess for differences between device measures of dorsal mobility. Intraclass correlation coefficient (ICC) and absolute difference values were computed to further assess the agreement in measures.
RESULTS: Dorsal mobility measured with the Glasoe device averaged 4.9 mm (1.8 to 9.3 mm). Dorsal mobility measured with the Klaue device averaged 5.2 mm (2.5 to 8.5 mm). Paired t-tests (p = 0.12) revealed no significant difference in measures. An ICC of 0.70 and a mean absolute difference of 0.9 mm (SD 0.8) were found between the two clinical measures further suggesting agreement.
CONCLUSION: Results indicated that the two devices possess similar diagnostic accuracy in the measurement of dorsal first ray mobility.

Entities:  

Mesh:

Year:  2005        PMID: 16309611     DOI: 10.1177/107110070502601111

Source DB:  PubMed          Journal:  Foot Ankle Int        ISSN: 1071-1007            Impact factor:   2.827


  8 in total

Review 1.  [Lapidus arthrodesis].

Authors:  C Plaaß; L Claaßen; S Ettinger; K Daniilidis; C Stukenborg-Colsman
Journal:  Orthopade       Date:  2017-05       Impact factor: 1.087

2.  Mobility of the first metatarsal-cuneiform joint in patients with and without hallux valgus: in vivo three-dimensional analysis using computerized tomography scan.

Authors:  Xiang Geng; Chen Wang; Xin Ma; Xu Wang; Jiazhang Huang; Chao Zhang; Jian Xu; Junsheng Yang
Journal:  J Orthop Surg Res       Date:  2015-09-15       Impact factor: 2.359

Review 3.  Hallux valgus and hypermobility of the first ray: facts and fiction.

Authors:  Jesse F Doty; Michael J Coughlin
Journal:  Int Orthop       Date:  2013-07-30       Impact factor: 3.075

Review 4.  Musculoskeletal conditions of the foot and ankle: assessments and treatment options.

Authors:  Smita Rao; Jody L Riskowski; Marian T Hannan
Journal:  Best Pract Res Clin Rheumatol       Date:  2012-06       Impact factor: 4.098

5.  Investigation of first ray mobility during gait by kinematic fluoroscopic imaging--a novel method.

Authors:  Heiner Martin; Ulf Bahlke; Albrecht Dietze; Volker Zschorlich; Klaus-Peter Schmitz; Thomas Mittlmeier
Journal:  BMC Musculoskelet Disord       Date:  2012-02-09       Impact factor: 2.362

6.  Three-Dimensional Printed Anatomical Models Help in Correcting Foot Alignment in Hallux Valgus Deformities.

Authors:  Anil Murat Ozturk; Onur Suer; Istemihan Coban; Mehmet Asim Ozer; Figen Govsa
Journal:  Indian J Orthop       Date:  2020-04-23       Impact factor: 1.251

7.  Is the Planus Foot Type Associated With First Ray Hypermobility?

Authors:  Oliver J Morgan; Rajshree Hillstrom; Robert Turner; Jonathan Day; Ibadet Thaqi; Kristin Caolo; Scott Ellis; Jonathan T Deland; Howard J Hillstrom
Journal:  Foot Ankle Orthop       Date:  2022-03-04

8.  The Validity and Reliability of a New Simple Instrument for the Measurement of First Ray Mobility.

Authors:  Pedro V Munuera-Martínez; Priscila Távara-Vidalón; Manuel A Monge-Vera; Antonia Sáez-Díaz; Guillermo Lafuente-Sotillos
Journal:  Sensors (Basel)       Date:  2020-04-14       Impact factor: 3.576

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.