Effects of gas leak on triggering function, humidification, and inspiratory oxygen fraction during noninvasive positive airway pressure ventilation.

Eriko Miyoshi1, Yuji Fujino, Akinori Uchiyama, Takashi Mashimo, Masaji Nishimura.   

Abstract

OBJECTIVES: During noninvasive positive pressure ventilation (NPPV), the gas leak that commonly occurs around the mask can render NPPV ineffective. We evaluated the effects of gas leak on inspiratory trigger function during NPPV with bilevel pressure and ICU ventilators. In addition, we evaluated the effects of gas leak on fraction of inspired oxygen (Fio(2)) and humidification.
METHODS: Air leak was created at the airway opening of a model lung by establishing several different-size holes in the circuit. During simulated spontaneous breathing, we evaluated inspiratory trigger performance of two bilevel pressure ventilators (BiPAP Vision and BiPAP S/T-D; Respironics; Murrysville, PA) and two ICU ventilators (Puritan-Bennett 7200ae and Puritan-Bennett 840; Tyco Healthcare; Mansfield, MA). Inspiratory delay time and inspiratory trigger pressure were analyzed. Fio(2) at the airway opening and inside the model lung were evaluated during BiPAP S/T-D ventilation at supplemental oxygen flows of 3, 6, 9, 12 and 15 L/min. Measured oxygen concentration was compared to mathematically predicted levels. Finally, using two heated humidifiers, we evaluated the effect of gas leak on humidification.
RESULTS: The bilevel pressure ventilators triggered properly at all levels of gas leak, and inspiratory triggering was more effective than with the ICU ventilators. Delivered Fio(2) with the BiPAP S/T-D ventilator was affected by gas leak and could be predicted mathematically unless the gas leak was large. With large gas leaks, although relative humidity was maintained, absolute humidity decreased.
CONCLUSION: Gas leak affected triggering of ICU ventilators, Fio(2) of the BiPAP S/T-D ventilator, and humidity with both types of humidifiers.

Entities:  

Mesh:

Year:  2005        PMID: 16304335     DOI: 10.1378/chest.128.5.3691

Source DB:  PubMed          Journal:  Chest        ISSN: 0012-3692            Impact factor:   9.410


  14 in total

1.  Bench studies evaluating devices for non-invasive ventilation: critical analysis and future perspectives.

Authors:  Carlo Olivieri; Roberta Costa; Giorgio Conti; Paolo Navalesi
Journal:  Intensive Care Med       Date:  2011-11-29       Impact factor: 17.440

2.  Performance of noninvasive ventilation modes on ICU ventilators during pressure support: a bench model study.

Authors:  Laurence Vignaux; Didier Tassaux; Philippe Jolliet
Journal:  Intensive Care Med       Date:  2007-06-12       Impact factor: 17.440

3.  Best clinical practices for the sleep center adjustment of noninvasive positive pressure ventilation (NPPV) in stable chronic alveolar hypoventilation syndromes.

Authors:  Richard B Berry; Alejandro Chediak; Lee K Brown; Jonathan Finder; David Gozal; Conrad Iber; Clete A Kushida; Timothy Morgenthaler; James A Rowley; Sally L Davidson-Ward
Journal:  J Clin Sleep Med       Date:  2010-10-15       Impact factor: 4.062

4.  Non-invasive Ventilation: Effect of Vented and Non-vented Exhalation Systems on Inspiratory CO2 and O2 Concentrations, Ventilation, and Breathing Pattern.

Authors:  Karl-Josef Franke; Maik Schroeder; Ulrike Domanski; Barbara Dewald; Georg Nilius
Journal:  Lung       Date:  2022-03-04       Impact factor: 2.584

5.  Cephalic versus oronasal mask for noninvasive ventilation in acute hypercapnic respiratory failure.

Authors:  Antoine Cuvelier; Wilfried Pujol; Stéphanie Pramil; Luis Carlos Molano; Catherine Viacroze; Jean-François Muir
Journal:  Intensive Care Med       Date:  2008-10-15       Impact factor: 17.440

6.  Inductive plethysmography to control volume-targeted ventilation for leak compensation.

Authors:  Marie Claire Andrieu; Cedric Quentin; David Orlikowski; Gilbert Desmarais; Daniel Isabey; Bruno Louis; Frédéric Lofaso
Journal:  Intensive Care Med       Date:  2008-03-21       Impact factor: 17.440

7.  Monitoring Patient/Ventilator Interactions: Manufacturer's Perspective.

Authors:  Gerard Evers; Carl Van Loey
Journal:  Open Respir Med J       Date:  2009-03-12

8.  Application of dual mask for postoperative respiratory support in obstructive sleep apnea patient.

Authors:  Jahan Porhomayon; Gino Zadeii; Nader D Nader; George R Bancroft; Alireza Yarahamadi
Journal:  Case Rep Anesthesiol       Date:  2013-04-10

Review 9.  Non-invasive ventilation in exacerbations of COPD.

Authors:  Nicolino Ambrosino; Guido Vagheggini
Journal:  Int J Chron Obstruct Pulmon Dis       Date:  2007

10.  Leaks can dramatically decrease FiO2 on home ventilators: a bench study.

Authors:  Philippe Goutorbe; Erwan Daranda; Yves Asencio; Pierre Esnault; Bertrand Prunet; Julien Bordes; Bruno Palmier; Eric Meaudre
Journal:  BMC Res Notes       Date:  2013-07-21
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.