Literature DB >> 16299050

Performance characteristics of a novel tandem mass spectrometry assay for serum testosterone.

Mark M Kushnir1, Alan L Rockwood, William L Roberts, Elizabeth G Pattison, Ashley M Bunker, Robert L Fitzgerald, A Wayne Meikle.   

Abstract

BACKGROUND: Commercial immunoassays for testosterone (Te) may give inaccurate results for samples from women and children, leading to misdiagnosis and inappropriate treatment. We developed a sensitive and specific tandem mass spectrometric assay for measurement of Te at the concentrations encountered in women and children.
METHODS: Te was extracted with methyl tert-butyl ether from 100 microL of serum or plasma, derivatized to form an oxime, and reextracted by solid-phase extraction. Instrumental analysis was performed on an API 4000 HPLC tandem mass spectrometer in the multiple-reaction monitoring (MRM) mode. The MRM transitions (m/z) were 304-->124 and 304-->112 for Te and 307-->124 and 307-->112 for d3-Te.
RESULTS: Within- and between-run CVs were <12% and 7.9%, respectively. The limit of quantification was 0.0346 nmol/L (1 ng/dL). Reference intervals for sex hormone-binding globulin and total, free, and bioavailable Te were established for children of Tanner stages 1 through 5 and adult males and females.
CONCLUSIONS: The sensitivity and specificity of the method are adequate for analysis of Te in samples from women and children. The method requires small sample volumes, has adequate precision, and is not subject to interferences.

Entities:  

Mesh:

Substances:

Year:  2005        PMID: 16299050     DOI: 10.1373/clinchem.2005.052167

Source DB:  PubMed          Journal:  Clin Chem        ISSN: 0009-9147            Impact factor:   8.327


  26 in total

1.  Reexamination of testosterone, dihydrotestosterone, estradiol and estrone levels across the menstrual cycle and in postmenopausal women measured by liquid chromatography-tandem mass spectrometry.

Authors:  Micol S Rothman; Nichole E Carlson; Mei Xu; Christina Wang; Ronald Swerdloff; Paul Lee; Victor H H Goh; E Chester Ridgway; Margaret E Wierman
Journal:  Steroids       Date:  2010-11-09       Impact factor: 2.668

2.  Principles and applications of liquid chromatography-mass spectrometry in clinical biochemistry.

Authors:  James J Pitt
Journal:  Clin Biochem Rev       Date:  2009-02

Review 3.  DIAGNOSIS OF ENDOCRINE DISEASE: How reliable are free thyroid and total T3 hormone assays?

Authors:  Kerry J Welsh; Steven J Soldin
Journal:  Eur J Endocrinol       Date:  2016-12       Impact factor: 6.664

Review 4.  Principles and clinical applications of liquid chromatography - tandem mass spectrometry for the determination of adrenal and gonadal steroid hormones.

Authors:  A E Kulle; M Welzel; P-M Holterhus; F G Riepe
Journal:  J Endocrinol Invest       Date:  2011-07-05       Impact factor: 4.256

5.  Simultaneous measurement of total estradiol and testosterone in human serum by isotope dilution liquid chromatography tandem mass spectrometry.

Authors:  Hui Zhou; Yuesong Wang; Matthew Gatcombe; Jacob Farris; Julianne C Botelho; Samuel P Caudill; Hubert W Vesper
Journal:  Anal Bioanal Chem       Date:  2017-08-11       Impact factor: 4.142

6.  Reference ranges for testosterone in men generated using liquid chromatography tandem mass spectrometry in a community-based sample of healthy nonobese young men in the Framingham Heart Study and applied to three geographically distinct cohorts.

Authors:  Shalender Bhasin; Michael Pencina; Guneet Kaur Jasuja; Thomas G Travison; Andrea Coviello; Eric Orwoll; Patty Y Wang; Carrie Nielson; Frederick Wu; Abdelouahid Tajar; Fernand Labrie; Hubert Vesper; Anqi Zhang; Jagadish Ulloor; Ravinder Singh; Ralph D'Agostino; Ramachandran S Vasan
Journal:  J Clin Endocrinol Metab       Date:  2011-06-22       Impact factor: 5.958

7.  A nanocomposite consisting of graphene oxide, zeolite imidazolate framework 8, and a molecularly imprinted polymer for (multiple) fiber solid phase microextraction of sterol and steroid hormones prior to their quantitation by HPLC.

Authors:  Roya Mirzajani; Fatemeh Kardani; Zahra Ramezani
Journal:  Mikrochim Acta       Date:  2019-02-01       Impact factor: 5.833

8.  Revisiting hyper- and hypo-androgenism by tandem mass spectrometry.

Authors:  Flaminia Fanelli; Alessandra Gambineri; Marco Mezzullo; Valentina Vicennati; Carla Pelusi; Renato Pasquali; Uberto Pagotto
Journal:  Rev Endocr Metab Disord       Date:  2013-06       Impact factor: 6.514

9.  Validation of a testosterone and dihydrotestosterone liquid chromatography tandem mass spectrometry assay: Interference and comparison with established methods.

Authors:  Christina Wang; Steve Shiraishi; Andrew Leung; Sima Baravarian; Laura Hull; Victor Goh; Paul W N Lee; Ronald S Swerdloff
Journal:  Steroids       Date:  2008-05-21       Impact factor: 2.668

10.  Fusaric acid as a novel proton-affinitive derivatizing reagent for highly sensitive quantification of hydroxysteroids by LC-ESI-MS/MS.

Authors:  Kouwa Yamashita; Keiko Yamazaki; Sachiko Komatsu; Mitsuteru Numazawa
Journal:  J Am Soc Mass Spectrom       Date:  2009-11-14       Impact factor: 3.109