Literature DB >> 16286110

Vaginal mesh extrusion associated with use of Mentor transobturator sling.

Andrew L Siegel1.   

Abstract

OBJECTIVES: To describe my experience of vaginal mesh extrusion using the monofilament polypropylene transobturator sling, the Mentor ObTape. The Mentor ObTape was the first transobturator sling developed as an alternative to the retropubic commercially available suburethral slings for providing mid-urethral support as treatment of female stress urinary incontinence.
METHODS: Thirty patients underwent transobturator suburethral sling placement for anatomic stress urinary incontinence using the ObTape from October 2003 to January 2005. A retrospective chart review was performed to retrieve data on the safety, efficacy, complications, and outcomes using this product.
RESULTS: Six patients (20%) to date have presented with defective vaginal healing manifested by extrusion of the sling material. Five patients required surgical removal of the sling material, and one underwent a trial of conservative management. No urethral erosions were noted.
CONCLUSIONS: In my experience, the Mentor ObTape sling, which uses a nonwoven, minimally elastic, micropore, monofilament polypropylene mesh, incurs an unacceptably high rate of defective vaginal wound healing and mesh extrusion.

Entities:  

Mesh:

Year:  2005        PMID: 16286110     DOI: 10.1016/j.urology.2005.05.015

Source DB:  PubMed          Journal:  Urology        ISSN: 0090-4295            Impact factor:   2.649


  9 in total

Review 1.  Complications of anterior compartment vaginal surgery.

Authors:  Eric S Rovner
Journal:  Curr Urol Rep       Date:  2007-09       Impact factor: 3.092

2.  Mesh infection without erosion after ObTape sling insertion: a diagnostic challenge.

Authors:  Sharif I M F Ismail
Journal:  Int Urogynecol J Pelvic Floor Dysfunct       Date:  2007-02-28

Review 3.  Safety considerations for synthetic sling surgery.

Authors:  Jerry G Blaivas; Rajveer S Purohit; Matthew S Benedon; Gabriel Mekel; Michael Stern; Mubashir Billah; Kola Olugbade; Robert Bendavid; Vladimir Iakovlev
Journal:  Nat Rev Urol       Date:  2015-08-18       Impact factor: 14.432

4.  Mechanical properties of polypropylene mesh used in pelvic floor repair.

Authors:  J S Afonso; P A L S Martins; M J B C Girao; R M Natal Jorge; A J M Ferreira; T Mascarenhas; A A Fernandes; J Bernardes; E C Baracat; G Rodrigues de Lima; B Patricio
Journal:  Int Urogynecol J Pelvic Floor Dysfunct       Date:  2007-09-11

Review 5.  Biomechanical, Topological and Chemical Features That Influence the Implant Success of an Urogynecological Mesh: A Review.

Authors:  Carmelo De Maria; Vito Santoro; Giovanni Vozzi
Journal:  Biomed Res Int       Date:  2016-04-28       Impact factor: 3.411

6.  Complications associated with transobturator sling procedures: analysis of 233 consecutive cases with a 27 months follow-up.

Authors:  Isabelle Kaelin-Gambirasio; Sandrine Jacob; Michel Boulvain; Jean-Bernard Dubuisson; Patrick Dällenbach
Journal:  BMC Womens Health       Date:  2009-09-25       Impact factor: 2.809

7.  Transobturator tapes are preferable over transvaginal tapes for the management of female stress urinary incontinence: Against.

Authors:  Arun Chawla
Journal:  Indian J Urol       Date:  2009 Oct-Dec

8.  Complications of grafts used in female pelvic floor reconstruction: Mesh erosion and extrusion.

Authors:  Tanya M Nazemi; Kathleen C Kobashi
Journal:  Indian J Urol       Date:  2007-04

9.  3D Printing of Drug-Loaded Thermoplastic Polyurethane Meshes: A Potential Material for Soft Tissue Reinforcement in Vaginal Surgery.

Authors:  Juan Domínguez-Robles; Caterina Mancinelli; Elena Mancuso; Inmaculada García-Romero; Brendan F Gilmore; Luca Casettari; Eneko Larrañeta; Dimitrios A Lamprou
Journal:  Pharmaceutics       Date:  2020-01-13       Impact factor: 6.321

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.