Literature DB >> 16283846

Performance and safety of holmium: YAG laser optical fibers.

Bodo E Knudsen1, Randolph D Glickman, Kenneth J Stallman, Saher Maswadi, Ben H Chew, Darren T Beiko, John D Denstedt, Joel M H Teichman.   

Abstract

BACKGROUND AND
PURPOSE: Lower-pole ureteronephroscopy requires transmission of holmium:YAG energy along a deflected fiber. Current ureteroscopes are capable of high degrees of deflection, which may stress laser fibers beyond safe limits during lower-pole use. We hypothesized that optical fiber and safety measures differ among manufacturers.
MATERIALS AND METHODS: Small (200-273-microm) and medium-diameter (300-400-microm) Ho:YAG fibers were tested in a straight and 180 degrees bent configuration. Energy transmission was measured by an energy detector. Fiber durability was assessed by firing the laser in sequentially tighter bending diameters. The fibers were bent to 180 degrees with a diameter of 6 cm and run at 200- to 4000-mJ pulse energy to determine the minimum energy required to fracture the fiber. The bending diameter was decreased by 1-cm increments and testing repeated until a bending diameter of 1 cm was reached. The maximum deflection of the ACMI DUR-8E ureteroscope with each fiber in the working channel was recorded. The flow rate through the working channel of the DUR-8E was measured for each fiber.
RESULTS: The mean energy transmission differed among fibers (P < 0.001). The Lumenis SL 200 and the InnovaQuartz 400 were the best small and medium-diameter fibers, respectively, in resisting thermal breakdown (P < 0.01). The Dornier Lightguide Super 200 fractured repeatedly at a bend diameter of 2 cm and with the lowest energy (200 mJ). The other small fibers fractured only at a bend diameter of 1 cm. The Sharplan 200 and InnovaQuartz Sureflex 273T were the most flexible fibers, the Lumenis SL 365 the least. The flow rate was inversely proportional to four times the power of the diameter of the fiber.
CONCLUSIONS: Optical performance and safety differ among fibers. Fibers transmit various amounts of energy to their cladding when bent. During lower-pole nephroscopy with the fiber deflected, there is a risk of fiber fracture from thermal breakdown and laser-energy transmission to the endoscope. Some available laser fibers carry a risk of ureteroscope damage.

Entities:  

Mesh:

Substances:

Year:  2005        PMID: 16283846     DOI: 10.1089/end.2005.19.1092

Source DB:  PubMed          Journal:  J Endourol        ISSN: 0892-7790            Impact factor:   2.942


  8 in total

Review 1.  Holmium laser for stone management.

Authors:  Sean Pierre; Glenn M Preminger
Journal:  World J Urol       Date:  2007-03-06       Impact factor: 4.226

2.  Retained laser fiber in the nidus of a recurrent cystine urolith in an intact male English bulldog.

Authors:  Emmanuelle M Butty; Anne-Sophie Bua; Nick P Vanstone; Marilyn E Dunn
Journal:  Can Vet J       Date:  2019-01       Impact factor: 1.008

Review 3.  Optimizing use of the holmium:YAG laser for surgical management of urinary lithiasis.

Authors:  Abhishek P Patel; Bodo E Knudsen
Journal:  Curr Urol Rep       Date:  2014-04       Impact factor: 3.092

4.  A Comparative in Vitro Study of Power Output Deterioration over Time Between Ho:YAG Laser Fibers from Different Manufacturers as a Function of Deflection and Power Input.

Authors:  Andreas Bourdoumis; Panagiotis Christopoulos; Nirmal Raj; Artemis Fedder; Noor Buchholz
Journal:  Curr Urol       Date:  2016-02-10

Review 5.  Intracorporeal lithotripsy.

Authors:  Peter Alken
Journal:  Urolithiasis       Date:  2017-12-04       Impact factor: 3.436

6.  Update on ureteroscopy instrumentation.

Authors:  Renato N Pedro; Manoj Monga
Journal:  Indian J Urol       Date:  2010-07

7.  The cost analysis of flexible ureteroscopic lithotripsy in 302 cases.

Authors:  Cenk Gurbuz; Gokhan Atış; Ozgur Arikan; Ozgur Efilioglu; Asıf Yıldırım; Onur Danacıoglu; Turhan Caskurlu
Journal:  Urolithiasis       Date:  2013-12-10       Impact factor: 3.436

8.  Accessory instrumentation in flexible ureteroscopy: Evidence-based recommendation.

Authors:  Timothy Holden; Renato Nardi Pedro; Monoj Monga
Journal:  Indian J Urol       Date:  2008-10
  8 in total