A preliminary investigation of the imaging performance of photostimulable phosphor computed radiography using a new design of mammographic quality control test object.

A R Cowen1, D S Brettle, N J Coleman, G J Parkin.   

Abstract

Leeds Test Object TOR[MAM] has been designed to supplement the current FAXIL mammography test object TOR[MAX]. It contains a range of details that have a more natural radiographic appearance and has been designed as a test that more closely approximates the image quality achieved in clinical mammography. Physical aspects of the design and implementation of TOR[MAM] are presented. The TOR[MAM] has been used in a preliminary physical evaluation of the comparative image qualities produced by conventional (screen-film) and photostimulable phosphor computed mammography and the results are discussed. TOR[MAX] results are also presented. The influence of digital image processing (enhancement) on the image quality of computed mammograms is also considered. The results presented indicate the sensitivity of TOR[MAM].

Entities:  

Mesh:

Year:  1992        PMID: 1628185     DOI: 10.1259/0007-1285-65-774-528

Source DB:  PubMed          Journal:  Br J Radiol        ISSN: 0007-1285            Impact factor:   3.039


  3 in total

1.  Optimization of Image Quality and Dose in Digital Mammography.

Authors:  Agnes M F Fausto; M C Lopes; M C de Sousa; Tânia A C Furquim; Anderson W Mol; Fermin G Velasco
Journal:  J Digit Imaging       Date:  2017-04       Impact factor: 4.056

2.  Clinical aspects of direct digital mammography.

Authors:  G J Parkin
Journal:  J Digit Imaging       Date:  1995-02       Impact factor: 4.056

3.  3D-printed breast phantom for multi-purpose and multi-modality imaging.

Authors:  Yaoyao He; Yulin Liu; Brandon A Dyer; John M Boone; Shanshan Liu; Tiao Chen; Fenglian Zheng; Ye Zhu; Yong Sun; Yi Rong; Jianfeng Qiu
Journal:  Quant Imaging Med Surg       Date:  2019-01
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.