Comparison of full-field digital mammography to screen-film mammography with respect to contrast and spatial resolution in tissue equivalent breast phantoms.

Cherie M Kuzmiak1, Etta D Pisano, Elodia B Cole, Donglin Zeng, Charles B Burns, Craig Roberto, Dag Pavic, Yeonhee Lee, Bo Kyoung Seo, Marcia Koomen, David Washburn.   

Abstract

To determine if the improved contrast resolution of full-field digital mammography (FFDM) with reduced spatial resolution allows for superior or equal phantom object detection compared with screen-film mammography (SFM). Tissue equivalent breast phantoms simulating an adipose to glandular ratio of 50/50,30/70, and 20/80 were imaged according to each manufacturers' recommendation with four full-field digital mammography units (Fuji, Sectra, Fischer, and General Electric) and a screen-film mammography unit (MammoMatII 2000, Siemens, Munich, Germany). A total of 20 images were obtained in both hard- and soft-copy formats. For the purpose of soft-copy display, the screen-film hard-copy images were digitized with a 50 microm micron scanner. Six radiologists, experts in breast imaging, and three physicists, experts in scoring mammography phantoms, participated in a reader study where each reader scored each phantom for visibility of line-pairs and for 24 objects (fibers, clusters of specks, and masses). The data were recorded, entered into a database, and analyzed by a mixed-effect model. The limiting spatial resolution in line-pairs per millimeter visible with the digital units was less, regardless of display modality used, than that provided by the screen-film unit. The difference was statistically significant for the General Electric (p < 0.01) and Fuji digital mammography units (p = 0.03). With respect to the number of visible objects, a statistically significant higher number could be detected with the screen-film unit as compared to the Fischer (p < 0.01) and Sectra (p < 0.01) digital mammography units, but there was no significant difference between the other digital units and screen film. Overall, there was significantly better performance on the 50/50 phantom than with the 30/70 and 20/80 phantoms (p = 0.01, p < 0.01) for object visibility. For the digital mammography units, soft-copy display performed better than hard-copy display for the Fischer and Sectra images, but worse for Fuji and General Electric. In addition, soft-copy display of digitized screen-film images was significantly better than hard-copy display (p =0.02) of the original screen films for object visibility, but worse for spatial resolution. The higher contrast resolution of the FFDM units tested did not result in improved detection of line-pair resolution or objects in the phantoms tested versus screen-film mammography. The phantom performance of a digital mammography unit seems to be influenced by the type of detection task (line-pair resolution versus object visibility), the display modality (soft-copy versus hard-copy) chosen to score the phantoms, and the parenchymal pattern composition of the phantom.

Mesh:

Year:  2005        PMID: 16279068     DOI: 10.1118/1.2040710

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  4 in total

1.  Visualization of intracerebral arteries by synchrotron radiation microangiography.

Authors:  K Myojin; A Taguchi; K Umetani; K Fukushima; N Nishiura; T Matsuyama; H Kimura; D M Stern; Y Imai; H Mori
Journal:  AJNR Am J Neuroradiol       Date:  2007-05       Impact factor: 3.825

2.  Surgical mammography reporting in a limited resource environment.

Authors:  John P Mouton; Justus Apffelstaedt; Karin Baatjes
Journal:  World J Surg       Date:  2010-11       Impact factor: 3.352

3.  Mammographic density assessed on paired raw and processed digital images and on paired screen-film and digital images across three mammography systems.

Authors:  Anya Burton; Graham Byrnes; Jennifer Stone; Rulla M Tamimi; John Heine; Celine Vachon; Vahit Ozmen; Ana Pereira; Maria Luisa Garmendia; Christopher Scott; John H Hipwell; Caroline Dickens; Joachim Schüz; Mustafa Erkin Aribal; Kimberly Bertrand; Ava Kwong; Graham G Giles; John Hopper; Beatriz Pérez Gómez; Marina Pollán; Soo-Hwang Teo; Shivaani Mariapun; Nur Aishah Mohd Taib; Martín Lajous; Ruy Lopez-Riduara; Megan Rice; Isabelle Romieu; Anath Arzee Flugelman; Giske Ursin; Samera Qureshi; Huiyan Ma; Eunjung Lee; Reza Sirous; Mehri Sirous; Jong Won Lee; Jisun Kim; Dorria Salem; Rasha Kamal; Mikael Hartman; Hui Miao; Kee-Seng Chia; Chisato Nagata; Sudhir Vinayak; Rose Ndumia; Carla H van Gils; Johanna O P Wanders; Beata Peplonska; Agnieszka Bukowska; Steve Allen; Sarah Vinnicombe; Sue Moss; Anna M Chiarelli; Linda Linton; Gertraud Maskarinec; Martin J Yaffe; Norman F Boyd; Isabel Dos-Santos-Silva; Valerie A McCormack
Journal:  Breast Cancer Res       Date:  2016-12-19       Impact factor: 6.466

4.  Microradiography of microcalcifications in breast specimen: a new histological correlation procedure and the effect of improved resolution on diagnostic validity.

Authors:  H-J Langen; S Koehler; J Bielmeier; R Jocher; D Kranzfelder; N Jagusch; G Treutlein; Th Wetzler; J Müller; G Ott
Journal:  Radiol Res Pract       Date:  2012-10-11
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.