Literature DB >> 1626731

The Simon nitinol filter: evaluation by MR and ultrasound.

D Kim1, R R Edelman, C J Margolin, D H Porter, C R McArdle, B W Schlam, L E Gianturco, J B Siegel, M Simon.   

Abstract

In this prospective blinded study of inferior vena caval (IVC) patency, 18 patients underwent 25 duplex ultrasound (US) and magnetic resonance (MR) angiography examinations over an eight-month period following Simon nitinol filter placement. Clinical examination for lower extremity venous stasis and plain abdominal radiography were also performed. Twenty-three of 24 MR examinations and 11 of 24 US examinations were judged technically adequate by the blinded observers. One technically adequate US exam was false positive for intraluminal caval thrombus. Thirteen technically inadequate US examinations missed 3 complete caval occlusions and 2 partial occlusions. MR identified all patients with complete or partial caval occlusion. The authors conclude that duplex US reliably confirms IVC patency only when strict criteria for technical adequacy and interpretation are met (good visualization of filter and IVC above and below filter). MR, although expensive, more reliably identifies nonoccluding intraluminal thrombus and caval occlusion. It should be the noninvasive study of choice in symptomatic patients with venous stasis and patients with recurrent pulmonary emboli.

Entities:  

Mesh:

Substances:

Year:  1992        PMID: 1626731     DOI: 10.1177/000331979204300701

Source DB:  PubMed          Journal:  Angiology        ISSN: 0003-3197            Impact factor:   3.619


  4 in total

1.  ACCF/ACR/AHA/NASCI/SCMR 2010 expert consensus document on cardiovascular magnetic resonance: a report of the American College of Cardiology Foundation Task Force on Expert Consensus Documents.

Authors:  W Gregory Hundley; David A Bluemke; J Paul Finn; Scott D Flamm; Mark A Fogel; Matthias G Friedrich; Vincent B Ho; Michael Jerosch-Herold; Christopher M Kramer; Warren J Manning; Manesh Patel; Gerald M Pohost; Arthur E Stillman; Richard D White; Pamela K Woodard
Journal:  Circulation       Date:  2010-05-17       Impact factor: 29.690

Review 2.  ACCF/ACR/AHA/NASCI/SCMR 2010 expert consensus document on cardiovascular magnetic resonance: a report of the American College of Cardiology Foundation Task Force on Expert Consensus Documents.

Authors:  W Gregory Hundley; David A Bluemke; J Paul Finn; Scott D Flamm; Mark A Fogel; Matthias G Friedrich; Vincent B Ho; Michael Jerosch-Herold; Christopher M Kramer; Warren J Manning; Manesh Patel; Gerald M Pohost; Arthur E Stillman; Richard D White; Pamela K Woodard
Journal:  J Am Coll Cardiol       Date:  2010-06-08       Impact factor: 24.094

Review 3.  Evidence-Based Evaluation of Inferior Vena Cava Filter Complications Based on Filter Type.

Authors:  Steven E Deso; Ibrahim A Idakoji; William T Kuo
Journal:  Semin Intervent Radiol       Date:  2016-06       Impact factor: 1.513

4.  Feasibility and optimization of ultra-short echo time MRI for improved imaging of IVC-filters at 3.0 T.

Authors:  Gesine Knobloch; Scott Nagle; Timothy Colgan; Tilman Schubert; Kevin M Johnson; Peter Bannas; Geng Li; Louis Hinshaw; James Holmes; Scott B Reeder
Journal:  Abdom Radiol (NY)       Date:  2021-01
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.