Variation of DaTSCAN quantification between different gamma camera types.

Rosemary J Morton1, Matthew J Guy, Craig A Marshall, Elizabeth A Clarke, Paul J Hinton.   

Abstract

OBJECTIVE: To acquire data from a 123I filled Alderson phantom on different gamma cameras types and compare the relative uptake results from processing using the QuantiSPECT program (GE Healthcare).
METHODS: A DaTSCAN phantom was filled using the standard protocol and imaged on seven different gamma camera types and on two identical cameras of the same type. The standard GE Healthcare protocols for the given cameras were used. Aliquots of the striatum and brain background were counted in a gamma counter to determine variations in filling concentration. All the raw DaTSCAN SPECT data was imported into QuantiSPECT and processed by the three different algorithms (two box, three box and crescent) to determine the relative uptake in the striatum. Inter-operater and intra-operator variation was also determined.
RESULTS: The 10% variation in filling concentration found across the sites was compensated for in the final results. There was a 5-15% variation between cameras depending on the processing algorithm used. There was an intra-operator variation of between 5 and 12% which reflected the proportion of operator intervention within the processing method. There was no statistical variation between operators.
CONCLUSIONS: The transfer of a DaTSCAN database between camera types is feasible, but ideally all data would be acquired on a single camera type and phantom data used to normalize the database accordingly.

Mesh:

Substances:

Year:  2005        PMID: 16264362     DOI: 10.1097/00006231-200512000-00014

Source DB:  PubMed          Journal:  Nucl Med Commun        ISSN: 0143-3636            Impact factor:   1.690


  6 in total

Review 1.  SPECT imaging evaluation in movement disorders: far beyond visual assessment.

Authors:  Kosmas Badiavas; Elisavet Molyvda; Ioannis Iakovou; Magdalini Tsolaki; Kyriakos Psarrakos; Nikolaos Karatzas
Journal:  Eur J Nucl Med Mol Imaging       Date:  2010-12-02       Impact factor: 9.236

2.  The Role of CT-Based Attenuation Correction and Collimator Blurring Correction in Striatal Spect Quantification.

Authors:  J M Warwick; S Rubow; M du Toit; E Beetge; P Carey; P Dupont
Journal:  Int J Mol Imaging       Date:  2011-04-06

3.  Comparison of machine learning and semi-quantification algorithms for (I123)FP-CIT classification: the beginning of the end for semi-quantification?

Authors:  Jonathan Christopher Taylor; John Wesley Fenner
Journal:  EJNMMI Phys       Date:  2017-11-29

4.  A new visual rating scale for Ioflupane imaging in Lewy body disease.

Authors:  Jim J Lloyd; George Petrides; Paul C Donaghy; Sean J Colloby; Johannes Attems; John T O'Brien; Gemma Roberts; Alan J Thomas
Journal:  Neuroimage Clin       Date:  2018-09-20       Impact factor: 4.881

5.  The subresolution DaTSCAN phantom: a cost-effective, flexible alternative to traditional phantom technology.

Authors:  Jonathan C Taylor; Nicholas Vennart; Ian Negus; Robin Holmes; Oliver Bandmann; Christine Lo; John Fenner
Journal:  Nucl Med Commun       Date:  2018-03       Impact factor: 1.690

6.  Effects of 99mTc-TRODAT-1 drug template on image quantitative analysis.

Authors:  Cheng-Han Wu; Bang-Hung Yang; Yuan-Hwa Chou; Shyh-Jen Wang; Jyh-Cheng Chen
Journal:  PLoS One       Date:  2018-03-15       Impact factor: 3.240

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.