Interdependence between measures of extent and severity of myocardial perfusion defects provided by automatic quantification programs.

Henrik Hussein El-Ali1, John Palmer, Marcus Carlsson, Lars Edenbrandt, Michael Ljungberg.   

Abstract

AIM: To evaluate the accuracy of the values of lesion extent and severity provided by the two automatic quantification programs AutoQUANT and 4D-MSPECT using myocardial perfusion images generated by Monte Carlo simulation of a digital phantom. The combination between a realistic computer phantom and an accurate scintillation camera simulation tool allows the generation of realistic single-photon emission computed tomography (SPECT) images similar to those obtained in clinical patient studies.
METHODS: The NCAT phantom and the SIMIND Monte Carlo program were used to simulate myocardial perfusion studies. Perfusion defects with sizes ranging from 5 to 17% of the left ventricular wall volume and reductions in tracer uptake of 20, 60 and 100% were simulated in three vascular territories.
RESULTS: The values of the extent provided by the programs were dependent on the reduction in tracer uptake, i.e. the severity. Similarly, the measures of severity were dependent on the size of the lesions. The severity provided by AutoQUANT for different defects was not dependent on the location, whereas 4D-MSPECT presented different values depending on the location in the left ventricle. The measures of extent and severity of the defects with the same true extent and activity uptake reduction provided by the two programs were different.
CONCLUSIONS: The NCAT phantom and the SIMIND Monte Carlo program were shown to be useful in simulating clinical myocardial SPECT studies. The quantification programs gave values of lesion extent that were dependent on the magnitude of the severity. Users should therefore consider this dependence when interpreting results from these programs.

Entities:  

Mesh:

Year:  2005        PMID: 16264361     DOI: 10.1097/00006231-200512000-00013

Source DB:  PubMed          Journal:  Nucl Med Commun        ISSN: 0143-3636            Impact factor:   1.690


  2 in total

1.  Validation of threshold method for myocardial control database by use of clinical data.

Authors:  Atsushi Narita; Susumu Shiomi; Joji Kawabe; Hiroyuki Tsushima; Takashi Yamanaga; Yasuyoshi Watanabe
Journal:  Radiol Phys Technol       Date:  2014-06-06

2.  Diagnostic sensitivity of SPECT myocardial perfusion imaging using a pumping cardiac phantom with inserted variable defects.

Authors:  Isabelle Chrysanthou-Baustert; Yiannis Parpottas; Ourania Demetriadou; Stelios Christofides; Charalambos Yiannakkaras; Demetris Kaolis; Marta Wasilewska-Radwanska; Tomasz Fiutowski; Franciszek Sikora
Journal:  J Nucl Cardiol       Date:  2013-05-25       Impact factor: 5.952

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.