Fast and accurate measurements of T1 using a multi-readout single inversion-recovery sequence.

P A Gowland1, M O Leach.   

Abstract

TOMROP is a multiple readout single inversion-recovery sequence which may potentially allow the measurement of multi-exponential T1 recoveries in vivo by NMR imaging. It is shown that several important modifications must be made to this sequence to permit T1 to be measured accurately and reliably. With these modifications it is possible to measure T1 with a mean systematic error of -1.5% (-0.3% for samples with T1 greater than 300 ms), a mean accuracy of 3.1% (2.1% for samples with T1 greater than 300 ms) and a mean repeatability of 1.4% (1.4% for samples with T1 greater than 300 ms), using data collected in 6.4 min from sixteen 128 x 128 matrix images.

Mesh:

Year:  1992        PMID: 1625570     DOI: 10.1002/mrm.1910260109

Source DB:  PubMed          Journal:  Magn Reson Med        ISSN: 0740-3194            Impact factor:   4.668


  9 in total

Review 1.  Cardiac T(1) imaging.

Authors:  Michael Jerosch-Herold; Raymond Y Kwong
Journal:  Top Magn Reson Imaging       Date:  2014-02

2.  Optimizing MR Scan Design for Model-Based ${T}_{1}$ , ${T}_{2}$ Estimation From Steady-State Sequences.

Authors:  Gopal Nataraj; Jon-Fredrik Nielsen; Jeffrey A Fessler
Journal:  IEEE Trans Med Imaging       Date:  2016-10-04       Impact factor: 10.048

3.  An oxygen-consuming phantom simulating perfused tissue to explore oxygen dynamics and (19)F MRI oximetry.

Authors:  Steven H Ubert Baete; Jan Vandecasteele; Luc Colman; Wilfried De Neve; Yves De Deene
Journal:  MAGMA       Date:  2010-06-25       Impact factor: 2.310

4.  Delayed clearance of cerebrospinal fluid tracer from choroid plexus in idiopathic normal pressure hydrocephalus.

Authors:  Per Kristian Eide; Lars Magnus Valnes; Are Hugo Pripp; Kent-Andre Mardal; Geir Ringstad
Journal:  J Cereb Blood Flow Metab       Date:  2019-09-07       Impact factor: 6.200

5.  TOWERS: T-One with Enhanced Robustness and Speed.

Authors:  Cihat Eldeniz; Jürgen Finsterbusch; Weili Lin; Hongyu An
Journal:  Magn Reson Med       Date:  2015-07-31       Impact factor: 4.668

6.  Optimization of Look-Locker Turbo-Field Echo-Planar Imaging and Evaluation of Its Accuracy in Head and Neck 3D T1 Mapping.

Authors:  Masanori Maehara; Masahiko Monma; Takeshi Nitanai; Tetsuya Matsumoto; Yukiko Fukuma
Journal:  Magn Reson Med Sci       Date:  2015-12-28       Impact factor: 2.471

Review 7.  T1 Mapping for Myocardial Fibrosis by Cardiac Magnetic Resonance Relaxometry-A Comprehensive Technical Review.

Authors:  Christian R Hamilton-Craig; Mark W Strudwick; Graham J Galloway
Journal:  Front Cardiovasc Med       Date:  2017-03-16

8.  Optimizing the magnetization-prepared rapid gradient-echo (MP-RAGE) sequence.

Authors:  Jinghua Wang; Lili He; Hairong Zheng; Zhong-Lin Lu
Journal:  PLoS One       Date:  2014-05-30       Impact factor: 3.240

9.  Development and validation of a new MRI simulation technique that can reliably estimate optimal in vivo scanning parameters in a glioblastoma murine model.

Authors:  Andrea Protti; Kristen L Jones; Dennis M Bonal; Lei Qin; Letterio S Politi; Sasha Kravets; Quang-Dé Nguyen; Annick D Van den Abbeele
Journal:  PLoS One       Date:  2018-07-23       Impact factor: 3.240

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.