An evaluation of home volume ventilators that support open-circuit mouthpiece ventilation.

Louis J Boitano1, Joshua O Benditt.   

Abstract

BACKGROUND: Open-circuit mouthpiece ventilation (MPV) is a form of noninvasive ventilation that can be used to provide portable daytime ventilatory support for neuromuscular patients with chronic respiratory failure. MPV has been reported to reduce the risk of respiratory infection due to tracheostomy, and to improve cough and voice function and patient quality of life. Despite these potential benefits, mouthpiece ventilation is not widely used. This may be due in part to the fact that little information is available as to which ventilators can support this application.
OBJECTIVE: To determine which volume-cycled portable home ventilators currently available in the United States will support MPV, and what peak inspiratory flow rates create adequate circuit pressure to prevent low-pressure alarming.
METHODS: We used a commercially available MPV breathing circuit with a set tidal volume range of 500\N1,000 mL with each of 8 ventilators currently available in the United States.
RESULTS: Six of the 8 ventilators supported MPV: Respironics Lifecare PLV-100 and PLV Continuum, Mallinckrodt Achieva PSO2, Pulmonetics LTV800, Newport HT50, and Uni-Vent Eagle 754.

Entities:  

Mesh:

Year:  2005        PMID: 16253153

Source DB:  PubMed          Journal:  Respir Care        ISSN: 0020-1324            Impact factor:   2.258


  6 in total

Review 1.  Mouthpiece ventilation and complementary techniques in patients with neuromuscular disease: A brief clinical review and update.

Authors:  Tiago Pinto; Michelle Chatwin; Paolo Banfi; Joao Carlos Winck; Antonello Nicolini
Journal:  Chron Respir Dis       Date:  2017-02-24       Impact factor: 2.444

2.  Tidal volume during mouthpiece non-invasive home ventilation: When the choice is the right answer.

Authors:  Giuseppe Fiorentino; Antonio M Esquinas
Journal:  Chron Respir Dis       Date:  2016-08-20       Impact factor: 2.444

3.  Accuracy of tidal volume delivered by home mechanical ventilation during mouthpiece ventilation: A bench evaluation.

Authors:  Adam Ogna; Helene Prigent; Line Falaize; Karl Leroux; Dante Santos; Isabelle Vaugier; David Orlikowski; Frederic Lofaso
Journal:  Chron Respir Dis       Date:  2016-07-08       Impact factor: 2.444

4.  Twenty-four hour noninvasive ventilation in Duchenne muscular dystrophy: a safe alternative to tracheostomy.

Authors:  Doug A McKim; Nadia Griller; Carole LeBlanc; Andrew Woolnough; Judy King
Journal:  Can Respir J       Date:  2013 Jan-Feb       Impact factor: 2.409

5.  Home monitoring of daytime mouthpiece ventilation effectiveness in patients with neuromuscular disease.

Authors:  Julie Nardi; Karl Leroux; David Orlikowski; Hélène Prigent; Frédéric Lofaso
Journal:  Chron Respir Dis       Date:  2015-12-23       Impact factor: 2.444

6.  252nd ENMC international workshop: Developing best practice guidelines for management of mouthpiece ventilation in neuromuscular disorders. March 6th to 8th 2020, Amsterdam, the Netherlands.

Authors:  Michelle Chatwin; Miguel Gonçalves; Jesus Gonzalez-Bermejo; Michel Toussaint
Journal:  Neuromuscul Disord       Date:  2020-07-22       Impact factor: 4.296

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.