Thin-film coupled fluid-solid analysis of flow through the Ahmed glaucoma drainage device.

Matthew S Stay1, Tingrui Pan, J David Brown, Babak Ziaie, Victor H Barocas.   

Abstract

The Ahmed glaucoma valve (AGV) is a popular glaucoma drainage device, allowing maintenance of normal intraocular pressure in patients with reduced trabecular outflow facility. The uniquely attractive feature of the AGV, in contrast to other available drainage devices, is its variable resistance in response to changes in flow rate. As a result of this variable resistance, the AGV maintains a pressure drop between 7 and 12 mm Hg for a wide range of aqueous humor flow rates. In this paper, we demonstrate that the nonlinear behavior of the AGV is a direct result of the flexibility of the valve material. Due to the thin geometry of the system, the leaflets of the AGV were modeled using the von Kármán plate theory coupled to a Reynolds lubrication theory model of the aqueous humor flow through the valve. The resulting two-dimensional coupled steady-state partial differential equation system was solved by the finite element method. The Poisson's ratio of the valve was set to 0.45, and the modulus was regressed to experimental data, giving a best-fit value 4.2 MPa. Simulation results compared favorably with previous experimental studies and our own pressure-drop/flow-rate data. For an in vitro flow of 1.6 microL/min, we calculated a pressure drop of 5.8 mm Hg and measured a pressure drop of 5.2 +/- 0.4 mm Hg. As flow rate was increased, pressure drop rose in a strongly sublinear fashion, with a flow rate of 20 microL/min giving a predicted pressure drop of only 10.9 mm Hg and a measured pressure drop of 10.5 +/- 1.1 mm Hg. The AGV model was then applied to simulate in vivo conditions. For an aqueous humor flow rate of 1.5-3.0 microL/min, the calculated pressure drops were 5.3 and 6.3 mm Hg.

Entities:  

Mesh:

Substances:

Year:  2005        PMID: 16248307     DOI: 10.1115/1.1993662

Source DB:  PubMed          Journal:  J Biomech Eng        ISSN: 0148-0731            Impact factor:   2.097


  4 in total

1.  Intraoperative testing of opening and closing pressure predicts risk of low intraocular pressure after Ahmed glaucoma valve implantation.

Authors:  F Bochmann; A Kipfer; J Tarantino; C Kaufmann; L Bachmann; M Thiel
Journal:  Eye (Lond)       Date:  2014-07-25       Impact factor: 3.775

Review 2.  Ocular Fluid Mechanics and Drug Delivery: A Review of Mathematical and Computational Models.

Authors:  Ajay Bhandari
Journal:  Pharm Res       Date:  2021-12-22       Impact factor: 4.200

3.  Evaluation of Changes in Refractive Errors Before and After Trabeculectomy and Shunt Surgery.

Authors:  Farzaneh Dehghanian Nasrabadi; Hassan Hashemi; Mohammad Aghazadeh Amiri; Ghasem Fakhraie; Alireza Akbarzadeh Baghban; Ebrahim Jafarzadehpour; Abbasali Yekta; Mehdi Khabazkhoob
Journal:  Clin Optom (Auckl)       Date:  2020-07-29

4.  Is Ahmed Glaucoma Valve Consistent in Performance?

Authors:  Nikhil S Choudhari; Swathi Vallabh Badakere; Ashutosh Richhariya; Sai Naga Sri Harsha Chittajallu; Sirisha Senthil; Chandra Sekhar Garudadri
Journal:  Transl Vis Sci Technol       Date:  2018-06-22       Impact factor: 3.283