Literature DB >> 16246685

Complications, limitations, and failures of capsule endoscopy: a review of 733 cases.

Emanuele Rondonotti1, Juan Manuel Herrerias, Marco Pennazio, Angel Caunedo, Miguel Mascarenhas-Saraiva, Roberto de Franchis.   

Abstract

BACKGROUND: Although a variety of technical and clinical problems of capsule endoscopy have been reported, their incidence and clinical importance are unknown. The objective was to evaluate the incidence and the type of such events.
METHODS: This was a retrospective analysis of 733 consecutive examinations at 4 large referral centers. The main outcome measurements were that the problems were classified as "technical," i.e., related to the functioning of the equipment, and "clinical," i.e., related to patient characteristics. For each type of event, the causes and the impact on the ability to reach a diagnosis were examined.
RESULTS: A total of 183 problems were recorded in 174 patients (23.7%): one problem occurred in 165 patients, two in 9 patients. In 8.46% of patients, technical limitations or failures, or clinical complications prevented or hampered diagnosis. Technical problems (e.g., gaps in the recordings, short duration of capsule batteries, failure of downloading) occurred in the early phase of capsule use in 8.6% of examinations and prevented or hampered diagnosis in 2.9%. Clinical problems (difficulty/inability to swallow the capsule, incomplete small-bowel examination) occurred in 16.4% of examinations and hampered or prevented diagnosis in 6.4%. Capsule retention that required surgical or endoscopic retrieval occurred in 1.9% of cases.
CONCLUSIONS: Technical problems were rare and hampered or prevented the diagnosis in a very small number of cases. The majority of clinical failures were related to an incomplete small-bowel examination.

Entities:  

Mesh:

Year:  2005        PMID: 16246685     DOI: 10.1016/j.gie.2005.05.002

Source DB:  PubMed          Journal:  Gastrointest Endosc        ISSN: 0016-5107            Impact factor:   9.427


  64 in total

1.  Tracheal aspiration of a capsule endoscope: not always a benign event.

Authors:  Clare Parker; Carolyn Davison; Simon Panter
Journal:  Dig Dis Sci       Date:  2012-04-24       Impact factor: 3.199

2.  Factors associated with incomplete small bowel capsule endoscopy studies.

Authors:  Mitchell M Lee; Andrew Jacques; Eric Lam; Ricky Kwok; Pardis Lakzadeh; Ajit Sandhar; Brandon Segal; Sigrid Svarta; Joanna Law; Robert Enns
Journal:  World J Gastroenterol       Date:  2010-11-14       Impact factor: 5.742

3.  Surveillance using capsule endoscopy is safe in post-colectomy patients with familial adenomatous polyposis: a prospective Japanese study.

Authors:  Minori Matsumoto; Takeshi Nakajima; Yasuo Kakugawa; Taku Sakamoto; Shiko Kuribayashi; Yosuke Otake; Takahisa Matsuda; Yukihide Kanemitsu; Hirokazu Taniguchi; Yutaka Saito
Journal:  Fam Cancer       Date:  2016-01       Impact factor: 2.375

4.  Patency Capsule Tolerability in School-Aged Children.

Authors:  Daisuke Tokuhara; Kenji Watanabe; Yuki Cho; Haruo Shintaku
Journal:  Digestion       Date:  2017-06-22       Impact factor: 3.216

Review 5.  Small bowel capsule endoscopy in 2007: indications, risks and limitations.

Authors:  Emanuele Rondonotti; Federica Villa; Chris J J Mulder; Maarten A J M Jacobs; Roberto de Franchis
Journal:  World J Gastroenterol       Date:  2007-12-14       Impact factor: 5.742

Review 6.  Imaging of the small bowel in Crohn's disease: a review of old and new techniques.

Authors:  Simone Saibeni; Emanuele Rondonotti; Andrea Iozzelli; Luisa Spina; Gian-Eugenio Tontini; Flaminia Cavallaro; Camilla Ciscato; Roberto de Franchis; Francesco Sardanelli; Maurizio Vecchi
Journal:  World J Gastroenterol       Date:  2007-06-28       Impact factor: 5.742

Review 7.  The role of capsule endoscopy in acute gastrointestinal bleeding.

Authors:  Moshe Nadler; Rami Eliakim
Journal:  Therap Adv Gastroenterol       Date:  2014-03       Impact factor: 4.409

8.  Clinical Review of Small-Bowel Endoscopic Imaging.

Authors:  Michael J Bartel; Mark E Stark; Frank J Lukens
Journal:  Gastroenterol Hepatol (N Y)       Date:  2014-11

Review 9.  Current role of capsule endoscopy in Crohn's disease.

Authors:  Marisol Luján-Sanchis; Laura Sanchis-Artero; Laura Larrey-Ruiz; Laura Peño-Muñoz; Paola Núñez-Martínez; Génesis Castillo-López; Lara González-González; Carlos Boix Clemente; Cecilia Albert Antequera; Ana Durá-Ayet; Javier Sempere-Garcia-Argüelles
Journal:  World J Gastrointest Endosc       Date:  2016-09-16

10.  Long-term outcome in patients with obscure gastrointestinal bleeding after negative capsule endoscopy.

Authors:  Seong-Joon Koh; Jong Pil Im; Ji Won Kim; Byeong Gwan Kim; Kook Lae Lee; Sang Gyun Kim; Joo Sung Kim; Hyun Chae Jung
Journal:  World J Gastroenterol       Date:  2013-03-14       Impact factor: 5.742

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.