Literature DB >> 16234691

Reduction of magnetic resonance imaging-related heating in deep brain stimulation leads using a lead management device.

Kenneth B Baker1, Jean Tkach, John D Hall, John A Nyenhuis, Frank G Shellock, Ali R Rezai.   

Abstract

OBJECTIVE: To evaluate the ability of a lead management device to reduce magnetic resonance imaging (MRI)-related heating of deep brain stimulation (DBS) leads and thereby to decrease the risks of exposing patients with these implants to MRI procedures.
METHODS: Experiments were performed using the Activa series (Medtronic, Inc., Minneapolis, MN) DBS systems in an in vitro, gelled-saline head and torso phantom. Temperature change was recorded using fluoroptic thermometry during MRI performed using a transmit-and-receive radiofrequency body coil at 1.5 T and a transmit-and-receive radiofrequency head coil at 3 T. A cranial model placed in the phantom was used to test a custom-designed burr hole device that permitted the placement of small-diameter, concentric loops around the burr hole at the DBS lead as it exited the cranium.
RESULTS: A total of 41 scans were performed, with absolute temperature changes ranging from 0.8 to 10.3 degrees C. Depending on the MRI system tested and the side of the phantom on which the hardware was placed, loop placement resulted in reductions in temperature rise of 41 to 74%. The effect was linearly related to the number of loops formed (P < 0.01) over the range tested (0-2.75 loops).
CONCLUSION: Small, concentric loops placed around the burr hole seem to reduce MRI-related heating for these implants. Although the mechanism is still not fully understood, a device such as that used in the present study could permit a wider range of clinical scanning sequences to be used at 1.5 and 3 T in patients with DBS implants, in addition to increasing the margin of safety for the patient.

Entities:  

Mesh:

Substances:

Year:  2005        PMID: 16234691     DOI: 10.1227/01.neu.0000176877.26994.0c

Source DB:  PubMed          Journal:  Neurosurgery        ISSN: 0148-396X            Impact factor:   4.654


  20 in total

1.  Analysis of the role of lead resistivity in specific absorption rate for deep brain stimulator leads at 3T MRI.

Authors:  Leonardo M Angelone; Jyrki Ahveninen; John W Belliveau; Giorgio Bonmassar
Journal:  IEEE Trans Med Imaging       Date:  2010-03-22       Impact factor: 10.048

2.  Construction and modeling of a reconfigurable MRI coil for lowering SAR in patients with deep brain stimulation implants.

Authors:  Laleh Golestanirad; Maria Ida Iacono; Boris Keil; Leonardo M Angelone; Giorgio Bonmassar; Michael D Fox; Todd Herrington; Elfar Adalsteinsson; Cristen LaPierre; Azma Mareyam; Lawrence L Wald
Journal:  Neuroimage       Date:  2016-12-21       Impact factor: 6.556

3.  Feasibility of using linearly polarized rotating birdcage transmitters and close-fitting receive arrays in MRI to reduce SAR in the vicinity of deep brain simulation implants.

Authors:  Laleh Golestanirad; Boris Keil; Leonardo M Angelone; Giorgio Bonmassar; Azma Mareyam; Lawrence L Wald
Journal:  Magn Reson Med       Date:  2016-04-05       Impact factor: 4.668

4.  Local SAR near deep brain stimulation (DBS) electrodes at 64 and 127 MHz: A simulation study of the effect of extracranial loops.

Authors:  Laleh Golestanirad; Leonardo M Angelone; Maria Ida Iacono; Husam Katnani; Lawrence L Wald; Giorgio Bonmassar
Journal:  Magn Reson Med       Date:  2016-10-31       Impact factor: 4.668

5.  Effect of the extracranial deep brain stimulation lead on radiofrequency heating at 9.4 Tesla (400.2 MHz).

Authors:  Devashish Shrivastava; Aviva Abosch; Timothy Hanson; Jinfeng Tian; Akshay Gupte; Paul A Iaizzo; J Thomas Vaughan
Journal:  J Magn Reson Imaging       Date:  2010-09       Impact factor: 4.813

6.  Reducing RF-induced Heating near Implanted Leads through High-Dielectric Capacitive Bleeding of Current (CBLOC).

Authors:  Laleh Golestanirad; Leonardo M Angelone; John Kirsch; Sean Downs; Boris Keil; Giorgio Bonmassar; Lawrence L Wald
Journal:  IEEE Trans Microw Theory Tech       Date:  2019-01-01       Impact factor: 3.599

Review 7.  Improving Safety of MRI in Patients with Deep Brain Stimulation Devices.

Authors:  Alexandre Boutet; Clement T Chow; Keshav Narang; Gavin J B Elias; Clemens Neudorfer; Jürgen Germann; Manish Ranjan; Aaron Loh; Alastair J Martin; Walter Kucharczyk; Christopher J Steele; Ileana Hancu; Ali R Rezai; Andres M Lozano
Journal:  Radiology       Date:  2020-06-23       Impact factor: 11.105

8.  Heating induced near deep brain stimulation lead electrodes during magnetic resonance imaging with a 3 T transceive volume head coil.

Authors:  Devashish Shrivastava; Aviva Abosch; John Hughes; Ute Goerke; Lance DelaBarre; Rachana Visaria; Noam Harel; J Thomas Vaughan
Journal:  Phys Med Biol       Date:  2012-08-15       Impact factor: 3.609

9.  Development and testing of implanted carbon electrodes for electromagnetic field mapping during neuromodulation.

Authors:  Neeta Ashok Kumar; Munish Chauhan; Sri Kirthi Kandala; Sung-Min Sohn; Rosalind J Sadleir
Journal:  Magn Reson Med       Date:  2020-04-16       Impact factor: 4.668

Review 10.  Deep brain stimulation for Parkinson disease: an expert consensus and review of key issues.

Authors:  Jeff M Bronstein; Michele Tagliati; Ron L Alterman; Andres M Lozano; Jens Volkmann; Alessandro Stefani; Fay B Horak; Michael S Okun; Kelly D Foote; Paul Krack; Rajesh Pahwa; Jaimie M Henderson; Marwan I Hariz; Roy A Bakay; Ali Rezai; William J Marks; Elena Moro; Jerrold L Vitek; Frances M Weaver; Robert E Gross; Mahlon R DeLong
Journal:  Arch Neurol       Date:  2010-10-11
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.