Portable gas analyser Cosmed K4b2 compared to a laboratory based mass spectrometer system.

L R Mc Naughton1, R Sherman, S Roberts, D J Bentley.   

Abstract

AIM: The purpose of this study was to systematically test the accuracy of an automated, portable, gas analysis system, the Cosmed K4b2 with a laboratory based mass spectrometer system, the Morgan EX670 across a number of gas and ventilation parameters.
METHODS: Eight subjects (mean+/-SE) age, 23.7+/-1.1 y, height, 1.78+/-0.01 m, mass, 74.4+/-2.1 kg performed a V.O2max test and a submaximal exercise test at 150, 200, 250 and 300 Watts (W), on an SRM cycle ergometer. The Morgan EX670 system and the K4b2 were randomly connected in series, using the same breath for the calculation of gas and ventilatory parameters.
RESULTS: The K4b2 system reads significantly higher than the Morgan EX 670 for both VO2 and V.CO2 at 250 (VO2/V.CO2: p<0.05, p<0.002), and 300 W (VO2/V.CO2: p<0.002, p<0.005). Unsystematic bias between the 2 analysers varies between 1% and 16% and systematic bias between 3% and 8%.
CONCLUSION: There are some significant unsystematic and systematic differences between these 2 systems and laboratories should endeavour to utilise either one or the other piece of equipment to test their subjects.

Entities:  

Mesh:

Substances:

Year:  2005        PMID: 16230983

Source DB:  PubMed          Journal:  J Sports Med Phys Fitness        ISSN: 0022-4707            Impact factor:   1.637


  6 in total

1.  Real-world walking economy: can laboratory equations predict field energy expenditure?

Authors:  Peter G Weyand; Lindsay W Ludlow; Jennifer J Nollkamper; Mark J Buller
Journal:  J Appl Physiol (1985)       Date:  2021-08-19

2.  Comparison of the Cosmed K4b(2) portable metabolic system in measuring steady-state walking energy expenditure.

Authors:  Jennifer A Schrack; Eleanor M Simonsick; Luigi Ferrucci
Journal:  PLoS One       Date:  2010-02-18       Impact factor: 3.240

3.  Test-retest reliability and minimum detectable change using the K4b2: oxygen consumption, gait efficiency, and heart rate for healthy adults during submaximal walking.

Authors:  Benjamin J Darter; Kelly M Rodriguez; Jason M Wilken
Journal:  Res Q Exerc Sport       Date:  2013-06       Impact factor: 2.500

4.  Accuracy of the Cosmed K5 portable calorimeter.

Authors:  Scott E Crouter; Samuel R LaMunion; Paul R Hibbing; Andrew S Kaplan; David R Bassett
Journal:  PLoS One       Date:  2019-12-16       Impact factor: 3.240

5.  Effects of a commercially available branched-chain amino acid-alanine-carbohydrate-based sports supplement on perceived exertion and performance in high intensity endurance cycling tests.

Authors:  Marco Gervasi; Davide Sisti; Stefano Amatori; Sabrina Donati Zeppa; Giosuè Annibalini; Giovanni Piccoli; Luciana Vallorani; Piero Benelli; Marco B L Rocchi; Elena Barbieri; Anna R Calavalle; Deborah Agostini; Carmela Fimognari; Vilberto Stocchi; Piero Sestili
Journal:  J Int Soc Sports Nutr       Date:  2020-01-20       Impact factor: 5.150

6.  Effects of carbohydrate and protein supplementation during resistance exercise on respiratory exchange ratio, blood glucose, and performance.

Authors:  David M Laurenson; Danielle Jane Dubé
Journal:  J Clin Transl Endocrinol       Date:  2014-11-13
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.