Literature DB >> 16230128

Complications of transvaginal silicone-coated polyester synthetic mesh sling.

F E Govier1, K C Kobashi, D D Kuznetsov, C Comiter, P Jones, S E Dakil, R James.   

Abstract

OBJECTIVES: To report a premarket multicenter trial to test the feasibility of a transvaginal silicone-coated polyester synthetic mesh sling in women with anatomic incontinence.
METHODS: Fifty-one patients in four centers underwent transvaginal placement of a silicone-coated polyester synthetic mesh sling (American Medical Systems) during an 8-month period. Of the 51 patients, 31 were part of a prospective institutional review board-approved feasibility trial in three centers funded by American Medical Systems (group 1) and 20 underwent implantation by a single surgeon and their data were retrospectively reviewed (group 2). The studies were done concomitantly, and all slings were fixed transvaginally with bone anchors. All patients in group 1 were followed up at 4 weeks, 6 months, and 1 year (as applicable) with repeat questionnaires, physical examinations, and pad tests.
RESULTS: In group 1, 20 patients completed 6 months of follow-up. Ten patients (32%) required a second surgical procedure at an average of 183 days (range 68 to 343) postoperatively. Eight patients (26%) had vaginal extrusion of the mesh, one (3%) required sling lysis, and one (3%) required sling removal because of infection. In group 2, 8 patients (40%) underwent sling removal for vaginal extrusion at a mean of 160 days (range 83 to 214).
CONCLUSIONS: Transvaginally placed silicone-coated mesh slings used for the treatment of urinary incontinence demonstrated an unacceptably high vaginal extrusion rate in this study. Once identified, this study was immediately terminated, and this product was not marketed for this application in the United States.

Entities:  

Mesh:

Substances:

Year:  2005        PMID: 16230128     DOI: 10.1016/j.urology.2005.04.027

Source DB:  PubMed          Journal:  Urology        ISSN: 0090-4295            Impact factor:   2.649


  3 in total

1.  Mesh-related chronic infections in silicone-coated polyester suburethral slings.

Authors:  Joseph K-S Lee; Gerard Agnew; Peter L Dwyer
Journal:  Int Urogynecol J       Date:  2010-09-30       Impact factor: 2.894

Review 2.  Safety considerations for synthetic sling surgery.

Authors:  Jerry G Blaivas; Rajveer S Purohit; Matthew S Benedon; Gabriel Mekel; Michael Stern; Mubashir Billah; Kola Olugbade; Robert Bendavid; Vladimir Iakovlev
Journal:  Nat Rev Urol       Date:  2015-08-18       Impact factor: 14.432

3.  Mesh complications in female pelvic floor reconstructive surgery and their management: A systematic review.

Authors:  Hemendra N Shah; Gopal H Badlani
Journal:  Indian J Urol       Date:  2012-04