Feasibility study of the use of frozen human sera in split-sample comparison of immunoassays with candidate reference measurement procedures for total thyroxine and total triiodothyronine measurements.

Linda M Thienpont1, Katleen Van Uytfanghe, John Marriott, Peter Stokes, Lothar Siekmann, Anja Kessler, David Bunk, Susan Tai.   

Abstract

BACKGROUND: Diagnostic manufacturers must ensure/document metrologically traceable assays. We report on a feasibility study of a split-sample comparison for that purpose. Processed, frozen single-donation sera, assigned target values by candidate reference measurement procedures (cRMPs), were used with immunoassays for total thyroxine (TT(4)) and triiodothyronine (TT(3)) as models.
METHODS: Two serum panels were quantified for TT(4) and TT(3) with validated cRMPs and measured in parallel with at least 14 immunoassays. The results were interpreted in terms of traceability of calibration (trueness) and of the individual measurement result (accuracy) by linear regression analysis and graphical representation against specifications. The commutability of the sera was investigated by parallel analysis of TT(4) in freshly collected but nonfiltered specimens.
RESULTS: The TT(4) (TT(3)) concentrations in the sera (according to the cRMPs) were 64-269 nmol/L (0.88-13.7 nmol/L). The method comparison showed that for TT(4), on average, the immunoassays produced results in agreement with the cRMPs, whereas for TT(3), results were typically higher. It also demonstrated a considerable between-assay divergence in traceability of calibration and accuracy. The evidence of noncommutability of the sera attributable to processing, however, indicates that the interpretation should be treated with caution.
CONCLUSIONS: Frozen sera can be used for documenting/validating traceability of total thyroid measurements. The way in which the sera are processed may jeopardize commutability, however, and therefore requires in-depth investigation.

Entities:  

Mesh:

Substances:

Year:  2005        PMID: 16223886     DOI: 10.1373/clinchem.2005.058180

Source DB:  PubMed          Journal:  Clin Chem        ISSN: 0009-9147            Impact factor:   8.327


  2 in total

1.  Reference materials and commutability.

Authors:  Hubert W Vesper; W Gregory Miller; Gary L Myers
Journal:  Clin Biochem Rev       Date:  2007-11

2.  Reference materials and reference measurement procedures: an overview from a national metrology institute.

Authors:  David M Bunk
Journal:  Clin Biochem Rev       Date:  2007-11
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.