Literature DB >> 16212299

Incidence of complications during flap creation in LASIK using the NIDEK MK-2000 microkeratome in 26,600 cases.

Carlos Carrillo1, Arturo S Chayet, Paul J Dougherty, Miguel Montes, Roberto Magallanes, Jacobo Najman, Jacobo Fleitman, Alfredo Morales.   

Abstract

PURPOSE: To evaluate the incidence of intraoperative complications using the NIDEK MK-2000 microkeratome during flap creation in LASIK.
METHODS: The incidence of intraoperative flap complications during LASIK using the NIDEK MK-2000 microkeratome was retrospectively studied in 26,600 procedures performed in 5 outpatient excimer laser surgery centers.
RESULTS: A total of 65 (0.244%) complications were identified: 23 (0.086%) eyes had free caps, 13 (0.049%) eyes had an incomplete pass, 13 (0.049%) eyes had an epithelial defect, 11 (0.041%) eyes had buttonhole, and 5 (0.019%) eyes had irregular flap. The remaining 26,535 (99.756%) eyes had uneventful flap creation.
CONCLUSIONS: Use of the NIDEK MK-2000 microkeratome resulted in a low incidence of intraoperative complications, making it a safe and reliable device for creating flaps during LASIK.

Entities:  

Mesh:

Year:  2005        PMID: 16212299     DOI: 10.3928/1081-597X-20050902-20

Source DB:  PubMed          Journal:  J Refract Surg        ISSN: 1081-597X            Impact factor:   3.573


  8 in total

Review 1.  [Complications of corneal lamellar refractive surgery].

Authors:  T Kohnen; M Remy
Journal:  Ophthalmologe       Date:  2015-12       Impact factor: 1.059

2.  [Avoidance and management of complications in laser in situ keratomileusis].

Authors:  T Kohnen; O K Klaproth
Journal:  Ophthalmologe       Date:  2013-07       Impact factor: 1.059

3.  Incidence and management of epithelial loosening after LASIK.

Authors:  Faik Oruçoğlu; Mahir Kenduşim; Buket Ayoglu; Berrin Toksu; Sinan Goker
Journal:  Int Ophthalmol       Date:  2012-04-05       Impact factor: 2.031

4.  Intraoperative and early postoperative flap-related complications of laser in situ keratomileusis using two types of Moria microkeratomes.

Authors:  Yunus Karabela; Orkun Muftuoglu; Ibrahim Gokhan Gulkilik; Mehmet Selim Kocabora; Mustafa Ozsutcu
Journal:  Int Ophthalmol       Date:  2014-02-17       Impact factor: 2.031

5.  Intraoperative flap complications in laser in situ keratomileusis with two types of microkeratomes.

Authors:  Hani S Al-Mezaine; Saleh A Al-Amro; Saleh Al-Obeidan
Journal:  Saudi J Ophthalmol       Date:  2011-04-23

6.  Outcomes of Retreatment after Aborted Laser In Situ Keratomileusis due to Flap Complications.

Authors:  Hani S Al-Mezaine; Saleh A Al-Amro; Abdulaziz Al-Fadda; Saleh Al-Obeidan
Journal:  Middle East Afr J Ophthalmol       Date:  2011-07

7.  Corneal flap thickness with the Moria M2 single-use head 90 microkeratome in 72 consecutive LASIK procedures.

Authors:  Yunus Karabela; Orkun Muftuoglu; Faruk Kaya
Journal:  Clin Ophthalmol       Date:  2017-03-03

8.  Corneal aberrations after small-incision lenticule extraction versus Q value-guided laser-assisted in situ keratomileusis.

Authors:  Jun Zhang; Li Zheng; Xia Zhao; Yi Sun; Wei Feng; Minhui Yuan
Journal:  Medicine (Baltimore)       Date:  2019-02       Impact factor: 1.817

  8 in total