| Literature DB >> 16202668 |
Pengfei Li1, Yingwu Wang, Yan Wang, Yunbiao Tang, J Paul Fawcett, Yimin Cui, Jingkai Gu.
Abstract
A rapid, selective and sensitive method for the determination of the angiotensin II receptor antagonist, telmisartan, in human plasma has been developed. Telmisartan and the internal standard, diphenhydramine, were extracted from plasma using diethyl ether-dichloromethane (60:40, v/v), and separated on a Zorbax extend C(18) column using methanol-10mM ammonium acetate (85:15, v/v) adjusted to pH 4.5 after mixing with formic acid as mobile phase. Detection was carried out by multiple reaction monitoring on a Q-trap LC-MS/MS system with an ESI interface. The assay was linear over the range 0.5-600.0 ng/ml with a limit of quantitation of 0.5 ng/ml and a limit of detection of 0.05 ng/ml. Intra- and inter-day precision were <6.7% and <8.1%, respectively, and the accuracy was in the range 88.9-111.0%. The assay was applied to a pharmacokinetic study of telmisartan given as a single oral dose (80 mg) to healthy volunteers.Entities:
Mesh:
Substances:
Year: 2005 PMID: 16202668 DOI: 10.1016/j.jchromb.2005.09.021
Source DB: PubMed Journal: J Chromatogr B Analyt Technol Biomed Life Sci ISSN: 1570-0232 Impact factor: 3.205