Literature DB >> 16199626

Severe mesh complications following intravaginal slingplasty.

Kaven Baessler1, Alan D Hewson, Ralf Tunn, Bernhard Schuessler, Christopher F Maher.   

Abstract

OBJECTIVE: Synthetic meshes are increasingly used in the management of stress urinary incontinence and pelvic organ prolapse. This report describes severe complications following anterior and/or posterior intravaginal slingplasties employing a multifilament polypropylene mesh.
METHODS: We describe the symptoms, findings, subsequent management, and outcome of 19 consecutive women who have been referred with complications following anterior (n = 11) and/or posterior intravaginal slingplasty (n = 13) employing the multifilament polypropylene tape.
RESULTS: The main indications for removal of the 11 anterior intravaginal slings were intractable mesh infection in 6 women, retropubic abscess with cutaneous sinus in one, and vesico-vaginal fistula in one, intravesical mesh and pain syndrome in one, and voiding difficulties and pain syndrome in two. The main indications for removal of the 13 posterior intravaginal slings were intractable mesh infection in three and pain syndrome and dyspareunia in 10 women. Removal of the slings was performed after a median time of 24 months post-slingplasty. At follow-up between 6 weeks and 6 months, in all women genital pain, chronic vaginal discharge and bleeding, voiding, and defecation difficulties had been markedly alleviated (5) or they had ceased (14). Twelve of 17 sexually active women (71%) resumed sexual intercourse without difficulties. Ten women required subsequent surgery for stress incontinence and pelvic organ prolapse.
CONCLUSION: Surgeons should be aware of the potential complications of synthetic meshes. Until data on the safety and efficacy of the intravaginal slingplasties are available, these procedures cannot be recommended.

Entities:  

Mesh:

Substances:

Year:  2005        PMID: 16199626     DOI: 10.1097/01.AOG.0000177970.52037.0a

Source DB:  PubMed          Journal:  Obstet Gynecol        ISSN: 0029-7844            Impact factor:   7.661


  36 in total

1.  Clinical study of ischia spinous fascia fixation--a new pelvic reconstructive surgery.

Authors:  Lan Zhu; Jinghe Lang; Qingxia Zhang
Journal:  Int Urogynecol J       Date:  2011-04       Impact factor: 2.894

2.  A long-term review of posterior colporrhaphy with Vypro 2 mesh.

Authors:  Yik Nyok Lim; Reinhold Muller; Audrey Corstiaans; Sarah Hitchins; Christopher Barry; Ajay Rane
Journal:  Int Urogynecol J Pelvic Floor Dysfunct       Date:  2007-01-10

Review 3.  Xenograft use in reconstructive pelvic surgery: a review of the literature.

Authors:  Emanuel C Trabuco; Christopher J Klingele; John B Gebhart
Journal:  Int Urogynecol J Pelvic Floor Dysfunct       Date:  2007-01-17

Review 4.  The Integral Theory of continence.

Authors:  Peter E P Petros; Patrick J Woodman
Journal:  Int Urogynecol J Pelvic Floor Dysfunct       Date:  2007-10-30

5.  Gluteo-vaginal sinus formation complicating posterior intravaginal slingplasty followed by successful IVS removal. A case report and review of the literature.

Authors:  Themistoklis Mikos; Tryfon Tsalikis; Alexios Papanikolaou; Fotios Pournaropoulos; John N Bontis
Journal:  Int Urogynecol J Pelvic Floor Dysfunct       Date:  2007-09-21

Review 6.  Posterior compartment defect repair in vaginal surgery: update on surgical techniques.

Authors:  Van Anh T Ginger; Kathleen C Kobashi
Journal:  Curr Urol Rep       Date:  2007-09       Impact factor: 3.092

7.  Infected midurethral tape presenting as an ischiorectal abscess.

Authors:  Swati Jha; Stephen Radley; Andrew Shorthouse
Journal:  Int Urogynecol J Pelvic Floor Dysfunct       Date:  2008-06

8.  Multiple perineal abscesses and sinus tracts as a complication of vaginal mesh.

Authors:  Christina Lewicky-Gaupp; Edward J McGuire; Dee E Fenner
Journal:  Int Urogynecol J Pelvic Floor Dysfunct       Date:  2009-02-20

Review 9.  Complications of pelvic organ prolapse surgery and methods of prevention.

Authors:  Renaud de Tayrac; Loic Sentilhes
Journal:  Int Urogynecol J       Date:  2013-11       Impact factor: 2.894

10.  Perioperative complications in vaginal mesh procedures using trocar in pelvic organ prolapse repair.

Authors:  Fuat Demirci; Karakoc Birgul; Oya Demirci; Elif Demirci; Yavuz Akman; Erhan Karaalp; Nihal Dolgun
Journal:  J Obstet Gynaecol India       Date:  2013-05-15
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.