OBJECTIVE: This paper is focused on prototype development and accuracy evaluation of a medical Augmented Reality (AR) system. The accuracy of such a system is of critical importance for medical use, and is hence considered in detail. We analyze the individual error contributions and the system accuracy of the prototype.
MATERIALS AND METHODS: A passive articulated arm is used to track a calibrated end-effector-mounted video camera. The live video view is superimposed in real time with the synchronized graphical view of CT-derived segmented object(s) of interest within a phantom skull. The AR accuracy mostly depends on the accuracy of the tracking technology, the registration procedure, the camera calibration, and the image scanning device (e.g., a CT or MRI scanner).
RESULTS: The accuracy of the Microscribe arm was measured to be 0.87 mm. After mounting the camera on the tracking device, the AR accuracy was measured to be 2.74 mm on average (standard deviation = 0.81 mm). After using data from a 2-mm-thick CT scan, the AR error remained essentially the same at an average of 2.75 mm (standard deviation = 1.19 mm).
CONCLUSIONS: For neurosurgery, the acceptable error is approximately 2-3 mm, and our prototype approaches these accuracy requirements. The accuracy could be increased with a higher-fidelity tracking system and improved calibration and object registration. The design and methods of this prototype device can be extrapolated to current medical robotics (due to the kinematic similarity) and neuronavigation systems.

Mesh:

Year:  2005        PMID: 16199379     DOI: 10.3109/10929080500221626

Source DB:  PubMed          Journal:  Comput Aided Surg        ISSN: 1092-9088


  5 in total

1.  Comprehensive review of surgical microscopes: technology development and medical applications.

Authors:  Ling Ma; Baowei Fei
Journal:  J Biomed Opt       Date:  2021-01       Impact factor: 3.170

2.  Three-Dimensional Evaluation of Similarity of Right and Left Knee Joints.

Authors:  Ki-Mo Jang; Jong-Hoon Park; Minho Chang; Youngjun Kim; Deukhee Lee; Sehyung Park; Joon Ho Wang
Journal:  Knee Surg Relat Res       Date:  2017-12-01

3.  Augmented reality in healthcare education: an integrative review.

Authors:  Egui Zhu; Arash Hadadgar; Italo Masiello; Nabil Zary
Journal:  PeerJ       Date:  2014-07-08       Impact factor: 2.984

4.  Augmented reality in neurosurgery.

Authors:  Raniel Tagaytayan; Arpad Kelemen; Cecilia Sik-Lanyi
Journal:  Arch Med Sci       Date:  2016-03-22       Impact factor: 3.318

5.  Value of the surgeon's sightline on hologram registration and targeting in mixed reality.

Authors:  Javier A Luzon; Bojan V Stimec; Arne O Bakka; Bjørn Edwin; Dejan Ignjatovic
Journal:  Int J Comput Assist Radiol Surg       Date:  2020-09-28       Impact factor: 2.924

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.