Literature DB >> 16195675

A rare case of shunt malfunction attributable to a broken Codman-Hakim programmable shunt valve after a blow to the head.

T Okazaki1, S Oki, K Migita, K Kurisu.   

Abstract

We report a rare case of shunt malfunction resulting from breakage of the Codman-Hakim programmable shunt valve system following a blow to the head. Originally, the patient had suffered intraventricular hemorrhage a month after birth; ventriculoperitoneal shunt was performed at 7 months. After that, the shunt valve functioned well for 7 years. The patient presented with headaches after receiving a blow behind the right ear. The headache became exacerbated with activity and improved with rest. Damage to the shunt valve was identified during surgery, with a crack at the top of the valve housing. Symptoms resolved following a change of the complete shunt system. The possibility of damage to the shunt valve following a blow should be considered in shunt patients presenting with headaches. Copyright 2005 S. Karger AG, Basel.

Entities:  

Mesh:

Year:  2005        PMID: 16195675     DOI: 10.1159/000087481

Source DB:  PubMed          Journal:  Pediatr Neurosurg        ISSN: 1016-2291            Impact factor:   1.162


  3 in total

Review 1.  Italian guidelines on the assessment and management of pediatric head injury in the emergency department.

Authors:  Liviana Da Dalt; Niccolo' Parri; Angela Amigoni; Agostino Nocerino; Francesca Selmin; Renzo Manara; Paola Perretta; Maria Paola Vardeu; Silvia Bressan
Journal:  Ital J Pediatr       Date:  2018-01-15       Impact factor: 2.638

2.  Shunt Overdrainage Caused by Displacement of the Pressure Control Cam after Pressure Adjustment.

Authors:  Su-Ho Kim; Yoon-Soo Lee; Min-Seok Lee; Sang-Jun Suh; Jeong-Ho Lee; Dong-Gee Kang
Journal:  Korean J Neurotrauma       Date:  2016-10-31

3.  Shunt Valve Rupture in Ventriculoperitoneal Shunt Failure.

Authors:  Mustafa Güdük; Ahmet Akbaş; Mürüvvet Ayten Tüzünalp; Gürkan Berikol; Murat Şakir Ekşi
Journal:  World Neurosurg       Date:  2020-09-08       Impact factor: 2.104