Abstract

AIM: To investigate clinically significant analytical interference in point-of-care electrolyte analysis caused by contamination of blood specimens with hand disinfectant.
METHODS: Six different hand hygiene products were added separately to heparinised blood samples in varying amounts as contaminant. The contaminated samples were analysed by three different blood gas and electrolyte analysers for assessing interference on measured whole blood sodium and potassium concentrations.
RESULTS: There were significant analytical interferences caused by hand hygiene product contamination that varied depending on the combination of disinfectant and analyser. Small amounts of Microshield Antibacterial Hand Gel contamination caused large increases in measured sodium concentration. Such effect was much greater compared with the other five products tested, and started to occur at much lower levels of contamination. There was a trend towards lower sodium results in blood samples contaminated with Hexol Antiseptic Lotion (Hexol), the hand hygiene product that we used initially. Apart from AiE Hand Sanitizer, all the other hand disinfectants, especially Hexol, significantly elevated the measured potassium concentration, particularly when a direct ion-selective electrode method was used for measurement.
CONCLUSIONS: Hand disinfectant products can significantly interfere with blood electrolyte analysis. Proper precautions must be taken against contamination since the resultant errors can adversely affect the clinical management of patients.

Entities:  

Mesh:

Substances:

Year:  2005        PMID: 16194829      PMCID: PMC7131148          DOI: 10.1080/00313020500169156

Source DB:  PubMed          Journal:  Pathology        ISSN: 0031-3025            Impact factor:   5.306


  10 in total

1.  The study of some possible measurement errors in clinical blood electrolyte potentiometric (ISE) analysers.

Authors:  Vlatko Rumenjak; Stjepan Milardović; Ivan Kruhak; Bozidar S Grabarić
Journal:  Clin Chim Acta       Date:  2003-09       Impact factor: 3.786

2.  Discrepancies between sodium concentrations measured by the Kodak Ektachem 700 and by dilutional and direct ion-selective electrode analyzers.

Authors:  J A Stone; J R Moriguchi; D R Notto; P E Murphy; C J Dass; L M Wessels; E F Freier
Journal:  Clin Chem       Date:  1992-12       Impact factor: 8.327

3.  Protein interference with ion-selective electrode measurement depends on reference electrode composition and design.

Authors:  R B Payne; B M Buckley; K M Rawson
Journal:  Ann Clin Biochem       Date:  1991-01       Impact factor: 2.057

4.  Protein does not interfere with the ion-selective electrode determination of calcium, sodium or potassium ions.

Authors:  N E Saris
Journal:  J Clin Chem Clin Biochem       Date:  1988-02

5.  A major outbreak of severe acute respiratory syndrome in Hong Kong.

Authors:  Nelson Lee; David Hui; Alan Wu; Paul Chan; Peter Cameron; Gavin M Joynt; Anil Ahuja; Man Yee Yung; C B Leung; K F To; S F Lui; C C Szeto; Sydney Chung; Joseph J Y Sung
Journal:  N Engl J Med       Date:  2003-04-07       Impact factor: 91.245

6.  Severe acute respiratory syndrome: clinical and laboratory manifestations.

Authors:  Christopher W K Lam; Michael H M Chan; Chun K Wong
Journal:  Clin Biochem Rev       Date:  2004-05

7.  Infection control for SARS in a tertiary neonatal centre.

Authors:  P C Ng; K W So; T F Leung; F W T Cheng; D J Lyon; W Wong; K L Cheung; K S C Fung; C H Lee; A M Li; K L E Hon; C K Li; T F Fok
Journal:  Arch Dis Child Fetal Neonatal Ed       Date:  2003-09       Impact factor: 5.747

8.  Benzalkonium interference with test methods for potassium and sodium.

Authors:  T R Koch; J D Cook
Journal:  Clin Chem       Date:  1990-05       Impact factor: 8.327

9.  Combined use of alcohol hand rub and gloves reduces the incidence of late onset infection in very low birthweight infants.

Authors:  P C Ng; H L Wong; D J Lyon; K W So; F Liu; R K Y Lam; E Wong; A F B Cheng; T F Fok
Journal:  Arch Dis Child Fetal Neonatal Ed       Date:  2004-07       Impact factor: 5.747

10.  Clinical presentations and outcome of severe acute respiratory syndrome in children.

Authors:  K L E Hon; C W Leung; W T F Cheng; P K S Chan; W C W Chu; Y W Kwan; A M Li; N C Fong; P C Ng; M C Chiu; C K Li; J S Tam; T F Fok
Journal:  Lancet       Date:  2003-05-17       Impact factor: 79.321

  10 in total
  3 in total

1.  Pseudohyperkalemia in a Patient with T-Cell Acute Lymphoblastic Leukemia and Hyperleukocytosis.

Authors:  Madison Merritt; Hannah Kline; Sudha Garimella; Robert Seigler
Journal:  J Pediatr Intensive Care       Date:  2018-02-01

2.  Comparison of the point-of-care blood gas analyzer versus the laboratory auto-analyzer for the measurement of electrolytes.

Authors:  Anunaya Jain; Imron Subhan; Mahesh Joshi
Journal:  Int J Emerg Med       Date:  2009-02-24

3.  Errors in potassium measurement: a laboratory perspective for the clinician.

Authors:  Jaya R Asirvatham; Viju Moses; Loring Bjornson
Journal:  N Am J Med Sci       Date:  2013-04
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.