A comparison of radial artery blood pressure determination between the Vasotrac device and invasive arterial blood pressure monitoring in adolescents undergoing scoliosis surgery.

Mary E McCann1, David Hill, Kristin C Thomas, David Zurakowski, Peter C Laussen.   

Abstract

UNLABELLED: The Vasotrac device (Medwave, Arden Hills, MN) allows noninvasive measurement of arterial blood pressure (BP) and arterial waveform. We have previously demonstrated agreement between the Vasotrac and continuous intraarterial BP monitoring in children with a stable circulation after cardiac surgery in the cardiac intensive care unit. To assess this monitor during different physiologic conditions, we studied the Vasotrac in anesthetized adolescent children undergoing scoliosis surgery in the prone position, with or without controlled hypotension. Eleven children undergoing surgery for idiopathic scoliosis were enrolled in this study. The anesthetic consisted of primarily a nitrous oxide and narcotic technique with controlled hypotension obtained using IV labetalol. Data were analyzed using correlations, mean error, and Bland-Altman plots. Noninvasive BP measured by the Vasotrac correlated closely with intraarterial BP. Waveforms displayed by the two systems were qualitatively similar. Correlation between the two methods for systolic, diastolic and mean BP was r = 0.82, r = 0.83, and r = 0.90, respectively. We conclude that noninvasive BP measurement using the Vasotrac monitor enables near-continuous and reliable monitoring of BP during anesthesia in the prone position and pharmacologic-induced hypotension. IMPLICATIONS: The Vasotrac, a noninvasive arterial blood pressure monitor, performs well in the prone position in adolescent patients undergoing scoliosis surgery using a technique of mild-to-moderate induced hypotension.

Entities:  

Mesh:

Year:  2005        PMID: 16192506     DOI: 10.1213/01.ane.0000171231.29328.9f

Source DB:  PubMed          Journal:  Anesth Analg        ISSN: 0003-2999            Impact factor:   5.108


  4 in total

1.  Innovative continuous non-invasive cuffless blood pressure monitoring based on photoplethysmography technology.

Authors:  Juan C Ruiz-Rodríguez; Adolf Ruiz-Sanmartín; Vicent Ribas; Jesús Caballero; Alejandra García-Roche; Jordi Riera; Xavier Nuvials; Miriam de Nadal; Oriol de Sola-Morales; Joaquim Serra; Jordi Rello
Journal:  Intensive Care Med       Date:  2013-06-06       Impact factor: 17.440

Review 2.  Detecting vital signs with wearable wireless sensors.

Authors:  Tuba Yilmaz; Robert Foster; Yang Hao
Journal:  Sensors (Basel)       Date:  2010-12-02       Impact factor: 3.576

3.  Noninvasive techniques for blood pressure measurement are not a reliable alternative to direct measurement: a randomized crossover trial in ICU.

Authors:  Sara Ribezzo; Eleonora Spina; Stefano Di Bartolomeo; Gianfranco Sanson
Journal:  ScientificWorldJournal       Date:  2014-01-30

4.  Non-invasive monitoring of blood pressure using the Philips Intellivue MP50 monitor cannot replace invasive blood pressure techniques in surgery patients under general anesthesia.

Authors:  Xianghu Meng; Guanghui Zang; Longchang Fan; Lei Zheng; Jinzhen Dai; Xueren Wang; Wei Xia; Jihong Liu; Chuanhan Zhang
Journal:  Exp Ther Med       Date:  2013-05-17       Impact factor: 2.447

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.