Real-time detection of Helicobacter Pylori infection and atrophic gastritis: comparison between conventional methods and a novel device for gastric juice analysis during endoscopy.

A Tucci1, P Tucci, M Bisceglia, A Marchegiani, G Papadopoli, P Fusaroli, A Spada, M O Pistoletto, M O Pistoletto, M Cristino, L Poli, A Villani, M Bucci, M Marinelli, G Caletti.   

Abstract

BACKGROUND AND STUDY AIMS: Gastric juice may represent a valuable source of clinicopathological information if properly analyzed. We evaluated the reliability and clinical validity of data obtained using an innovative device (the "Mt 21-42") that analyzes gastric juice, thus allowing the identification of Helicobacter pylori infection and atrophic gastritis of the oxyntic mucosa during endoscopy.
METHODS: Validation studies were carried out to evaluate the measuring performance of the device. In addition, the H. pylori status and the presence of atrophic gastritis were assessed in 150 patients undergoing upper gastrointestinal endoscopy. In all these patients the Mt 21-42 device was used to assist endoscopy. Conventional tests (involving histology, urease testing, urea breath testing, anti- H. pylori IgG, serum gastrin, pepsinogen, intrinsic factor and parietal cells autoantibodies, vitamin B12, and folate) were also performed for comparison with the Mt 21-42 results.
RESULTS: The measuring performance of the Mt 21-42 was good; for pH, the relative percent error and the coefficient of variation were 1.9 % +/- 4.2 and 1.3 %, respectively, and for ammonium they were 0.1 % +/- 0.2 % and 2.1 %. For the detection of H. pylori infection, the sensitivity and specificity of the device (96.7 % and 94.3 %) were similar to those of the urea breath test (90.5 % and 93.3 %) and serology (87.1 % and 88.8 %), and higher than those of the urease test (78.6 % and 98.7 %; P < 0.01) and routine histology (94.3 % and 76.3 %; P < 0.05). When compared with the currently available standard methods, use of the Mt 21-42 was found to be the most sensitive technique for the detection of atrophy (94.7 % vs. 5.3 % - 47.4 %; P < 0.001); the device failed to detect the disease in only one case (5 %), whereas failure rates of 53 % - 95 % were reported with the conventional methods.
CONCLUSION: Atrophic gastritis of the oxyntic mucosa is a risky condition that often goes undetected in current clinical practice. The Mt 21-42 is an effective, useful, and desirable tool that may help to overcome this diagnostic limitation; it produces time and cost savings and also allows the detection of H. pylori infection.

Entities:  

Mesh:

Substances:

Year:  2005        PMID: 16189769     DOI: 10.1055/s-2005-870373

Source DB:  PubMed          Journal:  Endoscopy        ISSN: 0013-726X            Impact factor:   10.093


  2 in total

1.  Real-time diagnosis of H. pylori infection during endoscopy: Accuracy of an innovative tool (EndoFaster).

Authors:  Guido Costamagna; Angelo Zullo; Alessandra Bizzotto; Cristiano Spada; Cesare Hassan; Maria Elena Riccioni; Clelia Marmo; Giuseppe Strangio; Teresa Antonella Di Rienzo; Giovanni Cammarota; Antonio Gasbarrini; Alessandro Repici
Journal:  United European Gastroenterol J       Date:  2015-10-13       Impact factor: 4.623

2.  Diagnostic Accuracy of EndoFaster® and Narrow-Band Imaging Endoscopy in Patients with Impaired Gastric Acid Secretion: A Real-Time Prospective Study.

Authors:  M Cazzato; G Esposito; G Galli; E Pilozzi; E Lahner; V D Corleto; A Zullo; E Di Giulio; B Annibale
Journal:  Gastroenterol Res Pract       Date:  2021-03-20       Impact factor: 2.260

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.