Literature DB >> 16189687

Comparative surface examinations on corund blasted titanium implants and explants in total hip arthroplasty.

Alexander Schuh1, Wolfgang Uter, Werner Kachler, Jürgen Göske, Günther Zeiler, Christoph Lill.   

Abstract

INTRODUCTION: Blasting techniques using corundum to create rough surfaces are used in many different processes for the surface treatment of hip implants. Recent evidence points an association between residual particles and early loosenings in the sense of third-body wear. MATERIALS AND
METHOD: The surfaces of five unused, original-packaged CLS stems and ARR-Titan supporting rings, respectively, were compared with five explanted CLS stems and ARR-Titan supporting rings.
RESULTS: The surface of the ARR-Titan supporting rings was adhered to Al2O3 particles at 23.2+/-2.5% on the implants and 12.3+/-3.0% on the explants (P < 0.0001). For the CLS stem, the surface to which Al2O3 particles adhered was 16.4+/-2.2 % on the implants, whereas the surface to which Al2O3 particles adhered was 12.4+/-3.3% on the explants (P = 0.0275). DISCUSSION: The results of this study show that corundum particles can be found to cover a high percentage on rough-blasted titanium surfaces of ready-to-use devices. In contrast, this contamination is significantly less on the explants treated in an identical fashion prior to implantation.

Entities:  

Mesh:

Substances:

Year:  2005        PMID: 16189687     DOI: 10.1007/s00402-005-0050-0

Source DB:  PubMed          Journal:  Arch Orthop Trauma Surg        ISSN: 0936-8051            Impact factor:   3.067


  5 in total

1.  How is wettability of titanium surfaces influenced by their preparation and storage conditions?

Authors:  D Scharnweber; F Schlottig; S Oswald; K Becker; H Worch
Journal:  J Mater Sci Mater Med       Date:  2009-10-23       Impact factor: 3.896

2.  Fabrication of strongly attached hydroxyapatite coating on titanium by hydrothermal treatment of Ti-Zn-PO4 coated titanium in CaCl 2 solution.

Authors:  Alireza Valanezhad; Kanji Tsuru; Kunio Ishikawa
Journal:  J Mater Sci Mater Med       Date:  2015-07-16       Impact factor: 3.896

Review 3.  [The importance of wear couples for younger endoprosthesis patients].

Authors:  J Kircher; P Bergschmidt; R Bader; D Kluess; E Besser-Mahuzir; A Leder; W Mittelmeier
Journal:  Orthopade       Date:  2007-04       Impact factor: 1.087

4.  Mechanical failure of total hip arthroplasties and associated risk factors.

Authors:  Henrik C Bäcker; Chia H Wu; Arne Kienzle; Carsten Perka; Clemens Gwinner
Journal:  Arch Orthop Trauma Surg       Date:  2022-01-28       Impact factor: 3.067

5.  Biological Safety Evaluation and Surface Modification of Biocompatible Ti-15Zr-4Nb Alloy.

Authors:  Yoshimitsu Okazaki; Shin-Ichi Katsuda
Journal:  Materials (Basel)       Date:  2021-02-04       Impact factor: 3.623

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.