CT computer-optimized high-dose-rate brachytherapy with surface applicator technique for scar boost radiation after breast reconstruction surgery.

Alexandra J Stewart1, Desmond A O'Farrell, Jennifer R Bellon, Jorgen L Hansen, Catherine Duggan, Maria A Czerminska, Robert A Cormack, Phillip M Devlin.   

Abstract

PURPOSE: Immediate breast reconstruction has become increasingly prevalent after mastectomy for breast cancer. Postoperative scar boost radiation for the reconstructed breast presents many planning challenges due to the shape, size, and curvature of the scar. High-dose-rate (HDR) surface applicator brachytherapy is a novel and effective method of delivering scar boost radiation. Two cases, one with a saline implant and one with a transverse rectus abdominis musculocutaneous flap reconstruction, illustrate the method and advantages of HDR optimization of surface applicators. METHODS AND MATERIALS: For 2 patients a mold of the breast was made with Aquaplast sheets. A reproducible system was used for arm positioning. Skin fiducials, including tattoos from external beam planning, were matched to fiducials on the mold. HDR catheters were sited on the mold at 1cm intervals, with the central catheter situated along the scar. Topographically, both scars demonstrated extreme curvature in both craniocaudal and mediolateral directions. A CT computer-optimized HDR plan was developed, with the reference dose prescribed at the skin surface. The dosimetry was compared to single-field and matched-field electron plans.
RESULTS: This surface applicator technique provided a uniform skin dose of 100% to the entire clinical target volume (CTV) without hot spots in both patients. The patient position and surface applicator setup were consistently reproducible. The patients tolerated the treatment well with minimal skin erythema. In the single-field electron plan, skin dose was decreased to 50% at the periphery of the scar. Matching fields addressed this depth dose decrement, but resulted in large localized hot spots of more than 200% centrally in each field.
CONCLUSION: CT computer-optimized HDR surface applicator brachytherapy provided a reproducible homogeneous method of treating highly curved scars on the reconstructed breast. Electron beam treatment would result in longer and more complex treatments yet still provide a less homogeneous dose than this surface applicator technique.

Entities:  

Mesh:

Year:  2005        PMID: 16182223     DOI: 10.1016/j.brachy.2005.03.003

Source DB:  PubMed          Journal:  Brachytherapy        ISSN: 1538-4721            Impact factor:   2.362


  3 in total

1.  Malignant soft tissue sarcoma of the shoulder treated by surface mould brachytherapy boost in an adjuvant setting.

Authors:  Ashutosh Mukherji; Mourougan Sinnatamby
Journal:  J Contemp Brachytherapy       Date:  2017-04-26

2.  Impact of GBBS algorithm on post-mastectomy scar boost irradiation of breast using catheter flap.

Authors:  Mourougan Sinnatamby; Vijayaprabhu Neelakandan; Gunaseelan Karunanidhi; Saravanan Kandasamy; Seenisamy Ramapandian; Muniyappan Kannan; Elakiya Sampath
Journal:  J Contemp Brachytherapy       Date:  2021-05-24

3.  Preclinical assessment of comfort and secure fit of thermobrachytherapy surface applicator (TBSA) on volunteer subjects.

Authors:  Kavitha Arunachalam; Oana I Craciunescu; Edward J Markewitz; Paolo F Maccarini; Jaime L Schlorff; Paul R Stauffer
Journal:  J Appl Clin Med Phys       Date:  2012-09-06       Impact factor: 2.102

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.