Literature DB >> 16176526

Long-term performance of screw-in atrial pacing leads: a randomized comparison of J-shaped and straight leads.

David Luria1, David Bar-Lev, Osnat Gurevitz, Hava Granit, Zeev Rotstein, Michael Eldar, Michael Glikson.   

Abstract

OBJECTIVE: In this prospective, randomized, control study we compared performance of J-shaped (JL) and straight atrial leads (SL) with active (screw-in) fixation mechanism, over a 5-year follow-up period. SUBJECTS AND METHODS: A total of 208 consecutive patients were randomized for implantation with a Medtronic 4568 JL model (n = 103) versus a Medtronic 4068 SL model (n = 105), and all lead-related measurements and complications were recorded.
RESULTS: Lead-related complications occurred in JL and SL with similar frequency (6% and 9%, respectively, P = 0.45). Lead dislodgment occurred in 8 (7.8%) cases in SL versus none in JL (P = 0.004). Lead malfunction and excessive pacing thresholds without macrodislodgment occurred more frequently in JL (11; 10.7%) than in SL (4; 3.8%), P = 0.055. Lead material failure (insulation break) occurred only once in SL at the 5-year checkup. Most complications occurred during the first year after implantation. Other events occurred in both leads with similar frequency and included: death in 39 and 35, loss of follow-up in 8 and 6, and development of atrial fibrillation in 9 and 15 patients, respectively. Of those who completed follow-up, lead impedance and pacing threshold increased significantly in both groups. P wave decreased in SL but not in JL.
CONCLUSION: More dislocations occurred in SL, but were counterbalanced by more lead malfunctions and excessive pacing thresholds in the JL group. Both leads demonstrated minor deterioration in electrical measurements during follow-up, with only 1 case of lead material failure in SL. Overall, both groups demonstrated favorable lead performance throughout follow-up.

Entities:  

Mesh:

Year:  2005        PMID: 16176526     DOI: 10.1111/j.1540-8159.2005.00204.x

Source DB:  PubMed          Journal:  Pacing Clin Electrophysiol        ISSN: 0147-8389            Impact factor:   1.976


  2 in total

1.  Combining current of injury and P-wave sensing optimized right atrial active-fixation leads implantation.

Authors:  Meng-Rong Chen; Fei-Long Zhang; Wei-Wei Wang; Xue-Hai Chen; Jian-Hua Chen; Liang-Long Chen
Journal:  J Thorac Dis       Date:  2019-04       Impact factor: 2.895

2.  A long-term, prospective, cohort study on the performance of right ventricular pacing leads: comparison of active-fixation with passive-fixation leads.

Authors:  Lie Liu; Jiaojiao Tang; Hu Peng; Shulin Wu; Chunying Lin; Dongli Chen; Qianhuan Zhang; Yuanhong Liang; Silin Chen; Yan Chen; Huiqiang Wei
Journal:  Sci Rep       Date:  2015-01-07       Impact factor: 4.379

  2 in total