Literature DB >> 16168854

A comparison of electron beam dose calculation accuracy between treatment planning systems using either a pencil beam or a Monte Carlo algorithm.

George X Ding1, Joanna E Cygler, Christine W Yu, Nina I Kalach, G Daskalov.   

Abstract

PURPOSE: To present a comparison of the accuracy of two commercial electron beam treatment planning systems: one uses a Monte Carlo algorithm and the other uses a pencil beam model for dose calculations. METHODS AND MATERIALS: For the same inhomogeneous phantoms and incident beams, measured dose distributions are compared with those predicted by the commercial treatment planning systems at different source-to-surface distances (SSDs). The accuracy of the pencil beam system for monitor unit calculations is also tested at various SSDs. Beam energies of 6-20 MeV are used.
RESULTS: The pencil beam model shows some serious limitations in predicting hot and cold spots in inhomogeneous phantoms for small low- or high-density inhomogeneities, especially for low-energy electron beams, such as 9 MeV. Errors (>10%) are seen in predicting high- and low-dose variations for three-dimensional inhomogeneous phantoms. The Monte Carlo calculated results generally agree much better with measurements.
CONCLUSIONS: The accuracy of the pencil beam calculations is difficult to predict because it depends on both the inhomogeneity geometry and location. The pencil beam calculations using CADPLAN result in large errors in phantoms containing three-dimensional type inhomogeneities. The Monte Carlo method in Theraplan Plus dose calculation module is shown to be more robust in accurately predicting dose distributions and monitor units under the tested conditions.

Mesh:

Substances:

Year:  2005        PMID: 16168854     DOI: 10.1016/j.ijrobp.2005.06.016

Source DB:  PubMed          Journal:  Int J Radiat Oncol Biol Phys        ISSN: 0360-3016            Impact factor:   7.038


  11 in total

1.  Factors Associated With Radiation Pneumonitis in Patients Receiving Electron Boost Radiation for Breast-Conserving Therapy: A Retrospective Review.

Authors:  Wataru Kawakami; Shigeyuki Takamatsu; Masashi Taka; Kaname Ishii; Tetsu Nakaichi; Kazuaki Funamoto; Kunihiko Yokoyama
Journal:  Adv Radiat Oncol       Date:  2020-10-24

2.  Silicon diodes as an alternative to diamond detectors for depth dose curves and profile measurements of photon and electron radiation.

Authors:  Christian Scherf; Christiane Peter; Jussi Moog; Jörg Licher; Eugen Kara; Klemens Zink; Claus Rödel; Ulla Ramm
Journal:  Strahlenther Onkol       Date:  2009-08-04       Impact factor: 3.621

3.  Validation of an electron Monte Carlo dose calculation algorithm in the presence of heterogeneities using EGSnrc and radiochromic film measurements.

Authors:  Jean-François Aubry; Hugo Bouchard; Igor Bessières; Frédéric Lacroix
Journal:  J Appl Clin Med Phys       Date:  2011-11-15       Impact factor: 2.102

4.  The Study of Field Equivalence Determined by the Modeled Percentage Depth Dose in Electron Beam Radiation Therapy.

Authors:  You-Guo Ma; Yan-Shan Zhang; Yan-Cheng Ye; Jia-Ming Wu
Journal:  Biomed Res Int       Date:  2021-10-08       Impact factor: 3.411

5.  Radiation dosimetry effect evaluation of a carbon fiber couch on novel uRT-linac 506c accelerator.

Authors:  Dazhen Jiang; Zhen Cao; Yongchang Wei; Tingting Cao; Jiuling Shen; Conghua Xie; Yunfeng Zhou; Hui Liu; Jun Zhang
Journal:  Sci Rep       Date:  2021-06-29       Impact factor: 4.379

6.  Application of a dummy eye shield for electron treatment planning.

Authors:  Sei-Kwon Kang; Soah Park; Taejin Hwang; Kwang-Ho Cheong; Taejin Han; Haeyoung Kim; Me-Yeon Lee; Kyoung Ju Kim; Do Hoon Oh; Hoonsik Bae
Journal:  J Radiat Res       Date:  2012-08-21       Impact factor: 2.724

Review 7.  From cultured to uncultured genome sequences: metagenomics and modeling microbial ecosystems.

Authors:  Daniel R Garza; Bas E Dutilh
Journal:  Cell Mol Life Sci       Date:  2015-08-09       Impact factor: 9.261

8.  Evaluation of treatment plans using various treatment techniques for the radiotherapy of cutaneous Kaposi's sarcoma developed on the skin of feet.

Authors:  Jong Min Park; Il Han Kim; Sung-Joon Ye; Kyubo Kim
Journal:  J Appl Clin Med Phys       Date:  2014-11-08       Impact factor: 2.102

9.  Comprehensive evaluation and clinical implementation of commercially available Monte Carlo dose calculation algorithm.

Authors:  Aizhen Zhang; Ning Wen; Teamour Nurushev; Jay Burmeister; Indrin J Chetty
Journal:  J Appl Clin Med Phys       Date:  2013-03-04       Impact factor: 2.102

10.  Convolution-based modified Clarkson integration (CMCI) for electron cutout factor calculation.

Authors:  Jina Chang; Mu-Han Lin; Weiguo Lu; Mingli Chen; Steve Jiang
Journal:  J Appl Clin Med Phys       Date:  2018-02-03       Impact factor: 2.102

View more