Literature DB >> 16160297

Automatic measurement of skin wheals provoked by skin prick tests.

Michael Prinz1, Kornelia Vigl, Stefan Wöhrl.   

Abstract

Skin prick tests (SPT) represent the standard method for the diagnosis of type-1 allergies. The skin wheals provoked by the SPTs are considered positive above certain cut-off diameters, usually 3 mm. At present their size is mostly estimated by measuring the diameter with rulers. Since the shape of wheals usually is not circular, the measurement of diameters leads to imprecise results. Therefore, we developed an algorithm for precisely measuring the wheals' area in mm(2) automatically. The average deviation of the automatic measurement on a test set achieved with the developed algorithm was 6.9 %. Compared to the maximum standard deviation of 6.59 % when measuring the manually redrawn outlines, the automatic method works sufficiently well.

Entities:  

Mesh:

Year:  2005        PMID: 16160297

Source DB:  PubMed          Journal:  Stud Health Technol Inform        ISSN: 0926-9630


  2 in total

1.  Measurement and interpretation of skin prick test results.

Authors:  J P M van der Valk; R Gerth van Wijk; E Hoorn; L Groenendijk; I M Groenendijk; N W de Jong
Journal:  Clin Transl Allergy       Date:  2016-02-23       Impact factor: 5.871

2.  Robust automated reading of the skin prick test via 3D imaging and parametric surface fitting.

Authors:  Jesus Pineda; Raul Vargas; Lenny A Romero; Javier Marrugo; Jaime Meneses; Andres G Marrugo
Journal:  PLoS One       Date:  2019-10-21       Impact factor: 3.240

  2 in total