| Literature DB >> 16153624 |
Sara Baldelli1, Stefania Zenoni, Simona Merlini, Norberto Perico, Dario Cattaneo.
Abstract
In the clinical practice of organ transplantation everolimus (RAD) is used in combination with cyclosporine (CsA), the most common antirejection agent. Both drugs show a narrow therapeutic window, which requires strict monitoring of their blood concentration. Simple methods for simultaneous measurement of RAD and CsA concentration are needed. As we have recently developed an HPLC-UV assay for RAD determination, we decided to implement it to allow concomitant measurement of CsA. The within- and between-day coefficients of variation of the measurement were less than 12.1% for RAD and 9.8% for CsA. The within- and between-day inaccuracy of quality control samples were less than 9.7% for RAD and less than 4.9% for CsA. The method was found accurate and precise and useful for simultaneous therapeutic monitoring of the two drugs.Entities:
Mesh:
Substances:
Year: 2005 PMID: 16153624 DOI: 10.1016/j.cca.2005.07.019