Literature DB >> 16151059

Experience with the recovery filter as a retrievable inferior vena cava filter.

William J Grande1, Scott O Trerotola, Patrick M Reilly, Timothy W I Clark, Michael C Soulen, Aalpen Patel, Richard D Shlansky-Goldberg, Catherine M Tuite, Jeffrey A Solomon, Jeffrey I Mondschein, Mary Kate Fitzpatrick, S William Stavropoulos.   

Abstract

PURPOSE: This study evaluates clinical experience with the Recovery filter as a retrievable inferior vena cava (IVC) filter.
MATERIALS AND METHODS: One hundred seven Recovery filters were placed in 106 patients with an initial clinical indication for temporary caval filtration. Patients were followed up to assess filter efficacy, complications, eventual need for filter removal, time to retrieval, and ability to remove the filter.
RESULTS: The patient cohort consisted of 62 men and 44 women with a mean age of 48 years (range, 18-90 y). Mean implantation time was 165 days. Indications for filter placement in patients with deep vein thrombosis (DVT) and/or pulmonary embolism (PE) included contraindication to anticoagulation (n = 33), complications of anticoagulation (n = 8), poor cardiopulmonary reserve (n = 6), large clot burden (n = 3), and PE while receiving anticoagulation (n = 1). Indications for filter placement in patients without proven PE or DVT included immobility after trauma (n = 35); recent intracranial hemorrhage, neurosurgery, or brain tumor (n = 18); and other surgical or invasive procedure (n = 3). Three patients (2.8%) had symptomatic PE after placement of the Recovery filter. No caval thromboses were detected. No symptomatic filter migrations occurred. Recovery filter removal was attempted in 15 of 106 patients (14%) at a mean of 150 days after placement. The Recovery filter was successfully retrieved in 14 of 15 patients (93%); one removal was unsuccessful at 210 days after placement. Ninety-two filters (87%) currently remain in place.
CONCLUSIONS: Although all the filters were placed with the intention of being removed, a large percentage of filters were not retrieved. The Recovery filter was safe and effective in preventing PE when used as a retrievable IVC filter.

Entities:  

Mesh:

Substances:

Year:  2005        PMID: 16151059     DOI: 10.1097/01.RVI.0000171689.52536.FD

Source DB:  PubMed          Journal:  J Vasc Interv Radiol        ISSN: 1051-0443            Impact factor:   3.464


  18 in total

1.  Inferior Vena Cava Filters: Indications, Outcomes, and Evidence.

Authors:  Jennifer P Montgomery; John A Kaufman
Journal:  Curr Treat Options Cardiovasc Med       Date:  2015-09

2.  Retrieval of Recovery IVC Filter After 1,463-Day Implantation.

Authors:  Michael S Stecker; Alisa Suzuki; Jonathan D Gates
Journal:  Eur J Trauma Emerg Surg       Date:  2009-03-16       Impact factor: 3.693

3.  Inferior vena cava filtration in the management of venous thromboembolism: filtering the data.

Authors:  Christopher Molvar
Journal:  Semin Intervent Radiol       Date:  2012-09       Impact factor: 1.513

Review 4.  Optional inferior vena caval filters: where are we now?

Authors:  A N Keeling; T B Kinney; M J Lee
Journal:  Eur Radiol       Date:  2008-04-02       Impact factor: 5.315

5.  Complications of inferior vena caval filters.

Authors:  Thuong G Van Ha
Journal:  Semin Intervent Radiol       Date:  2006-06       Impact factor: 1.513

6.  Retrieval of a wall-embedded recovery inferior vena cava filter using rigid bronchoscopy forceps.

Authors:  Sarah B White; S William Stavropoulos
Journal:  Semin Intervent Radiol       Date:  2007-03       Impact factor: 1.513

Review 7.  Are too many inferior vena cava filters used? Controversial evidences in different clinical settings: a narrative review.

Authors:  Michele Dalla Vestra; Elisabetta Grolla; Luca Bonanni; Raffaele Pesavento
Journal:  Intern Emerg Med       Date:  2016-11-21       Impact factor: 3.397

8.  Ten-year experience of retrievable inferior vena cava filters in a tertiary referral center.

Authors:  George Tse; Trevor Cleveland; Stephen Goode
Journal:  Diagn Interv Radiol       Date:  2017 Mar-Apr       Impact factor: 2.630

9.  Rates and predictors of plans for inferior vena cava filter retrieval in hospitalized patients.

Authors:  John F Mission; Robert K Kerlan; Justin H Tan; Margaret C Fang
Journal:  J Gen Intern Med       Date:  2010-01-20       Impact factor: 5.128

Review 10.  Optional vena cava filters: indications, management, and results.

Authors:  Hanno Hoppe
Journal:  Dtsch Arztebl Int       Date:  2009-06-12       Impact factor: 5.594

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.