Confirmation of nasogastric tube position by pH testing.

S J Taylor1, R Clemente.   

Abstract

INTRODUCTION: In 2004, the Medicines and Healthcare products Regulatory Agency (MHRA) advised that nasogastric (NG) tube position should be confirmed using pH strips or paper. However, gastric pH is raised by the use of H2-blockers and proton-pump inhibitors (PPIs) potentially producing false negative pH tests resulting in delayed feeding. In addition, colorimetric differentiation using pH strips may be more prone to bias and inaccuracy than direct pH measurements largely used to establish the threshold.
METHOD: To quantify this problem a 1 day survey of all the patients requiring NG and nasointestinal (NI) feeding was undertaken, to establish the numbers of patients receiving H2-Blockers or PPIs, with or without a safe swallow and the methods currently being used to confirm tube positioning. A second observational study was performed to establish the accuracy of six pH strips available to NHS trusts against four unlabelled pH solutions.
RESULTS: Forty-two per cent of patients receiving NG feeding were on H2-blockers or PPIs, including 13% who had a safe swallow for acidic drinks that could be subsequently aspirated to confirm position. In the second study 'testers' correctly identified pH's 3, 4, 5 and 6 with Mackery-Nagel 0-6, BDH 0-6 and 0-14 strips but overestimated pH 4 as pH 5 with Johnson 0-11 paper, underestimated pH 6 as pH 5 with Pehanon 0-12 paper and with Litmus classified pH 3-5 as acid (all), but half also classified pH 6 as acid.
CONCLUSION: Theoretically 29% of NG tube positions could not be confirmed by pH testing because of the usage of PPIs or H2-blockers and lack of swallow. Some pH strips are either inaccurate or their result misinterpreted by staff. Large surveys and trials of the actual efficacy and accuracy of pH testing are required.

Entities:  

Mesh:

Substances:

Year:  2005        PMID: 16150133     DOI: 10.1111/j.1365-277X.2005.00635.x

Source DB:  PubMed          Journal:  J Hum Nutr Diet        ISSN: 0952-3871            Impact factor:   3.089


  12 in total

Review 1.  Naso-enteric Tube Placement: A Review of Methods to Confirm Tip Location, Global Applicability and Requirements.

Authors:  S A Milsom; J A Sweeting; H Sheahan; E Haemmerle; J A Windsor
Journal:  World J Surg       Date:  2015-09       Impact factor: 3.352

2.  The effects of oral fluid intake an hour before cesarean section on regurgitation incidence.

Authors:  Zohreh Ghorashi; Vahidreza Ashori; Fariba Aminzadeh; Mitra Mokhtari
Journal:  Iran J Nurs Midwifery Res       Date:  2014-07

3.  Comparison of four techniques of nasogastric tube insertion in anaesthetised, intubated patients: A randomized controlled trial.

Authors:  Mohan Chandra Mandal; Sujata Dolai; Santanu Ghosh; Pallab Kumar Mistri; Rajiv Roy; Sekhar Ranjan Basu; Sabyasachi Das
Journal:  Indian J Anaesth       Date:  2014 Nov-Dec

4.  Development and validation of a lipase nasogastric tube position test.

Authors:  Oliver Anderson; Reuben Carr; Merrilee Harbinson; George Bushra Hanna
Journal:  BMJ Open Gastroenterol       Date:  2016-03-02

5.  Novel Visual Nasogastric Tube Insertion System: A Feasibility and Efficiency Study in a Manikin.

Authors:  Qiaoya Li; Juan Xie; Jinxing Wu; Rui Guo; Wenwen Ma; Gang Xu; Min Yang; Huisheng Deng
Journal:  Gastroenterol Res Pract       Date:  2016-11-23       Impact factor: 2.260

6.  Diagnostic accuracy of a pH stick, modified to detect gastric lipase, to confirm the correct placement of nasogastric tubes.

Authors:  Anne M Rowat; Catriona Graham; Martin Dennis
Journal:  BMJ Open Gastroenterol       Date:  2018-07-31

7.  Blind bedside postpyloric placement of spiral tube as rescue therapy in critically ill patients: a prospective, tricentric, observational study.

Authors:  Bo Lv; Linhui Hu; Lifang Chen; Bei Hu; Yanlin Zhang; Heng Ye; Cheng Sun; Xiunong Zhang; Huilan Lan; Chunbo Chen
Journal:  Crit Care       Date:  2017-09-26       Impact factor: 9.097

8.  Study to determine the likely accuracy of pH testing to confirm nasogastric tube placement.

Authors:  Anne M Rowat; Catriona Graham; Martin Dennis
Journal:  BMJ Open Gastroenterol       Date:  2018-06-09

9.  Validation of the RightSpot™ device for determination of gastric pH during nasogastric tube placement.

Authors:  Charles R Lambert; David Varlotta; Marjorie Posey; Jadie L Heberlein; Janice M Shirley
Journal:  Int J Emerg Med       Date:  2013-07-17

10.  Usability study of pH strips for nasogastric tube placement.

Authors:  Simone Borsci; Peter Buckle; Jeremy Huddy; Zenia Alaestante; Zhifang Ni; George B Hanna
Journal:  PLoS One       Date:  2017-11-30       Impact factor: 3.240

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.